8 . T r o u b l e s h o o t i n g
8 . 2 : T r o u b l e s h o o t i n g A l a r m s C l a s s I V
INFORMATION ONLY ALARMS (Class III)
Problem
Operation Mode
Arterial Pressure
AutoPilot
Alarm
Operator
Low Helium Tank
AutoPilot
Pressure
Operator
AUTOMATIC ALERTS ALARMS (Class IV)
Problem
Operation Mode
Possible Late
AutoPilot
Deflation
Erratic Triggering
AutoPilot
No ECG signal
AutoPilot
available
Information Only Alarms
8-22
Possible Cause(s)
AP has fallen below
set limit
HE tank is empty
HE tank is OFF
Possible Cause(s)
Electromechanical delay is Check deflation timing. If
less than 100 msec
ECG connected from
bedside monitor. Signal
delay is longer than 35 msec 5 lead ECG cable.
> 3 lead switches within
1 minute and no AP signal
available
Noisy ECG signal
> 3 trigger switches
between AP and Pacer
within 1 minute
ECG signal is not available Check ECG connections.
but IABP is triggering on
AP or pacer signal
Corrective Actions
Check patient
hemodynamics. Check for
AP disconnect.
Change HE tank
Open HE tank
Corrective Actions
deflation timing is too late
and patient hemodynamics
are compromised, select
Operator mode and
manually adjust timing
Consider using direct
patient connection with
Check ECG signal quality.
Change ECG electrodes.
Change ECG lead. Adjust
Autogain or select Man gain.
Select Operator mode.
Check patient condition.
Select Operator mode.
Select appropriate
trigger mode.
Reconnect ECG cable or
leads. Attach another ECG
source from patient or
monitor.