6 . O p e r a t i n g P r o c e d u r e s
6 . 1 : O p e r a t i n g I n s t r u c t i o n s
ACTION
5. Connect IAB
6. Select and verify
7. Initiate Counterpulsation
8. Set timing
9. Check BPW
10. Press 1:1 assist
11. Verify alarms are ON
Summary of Start-up Instructions for Operator mode
6-4
OPERATOR MODE
DETAILS
• Volume automatically set to full volume
• Press Trigger Key
trigger mode
• Choose trigger mode
• Verify Trigger mode is displayed below HR
• Verify Heart Rate
• Verify white bands on ECG/AP
• Press Pump ON
• Automatic purge
• Set inflation and deflation timing using
• Select 1:2
• Perform timing assessment
• Verify IAB size comparing AUG to BPW
• Verify BPW shape
• Provides full assist
< and > timing keys
plateau pressure