6. Heat is detrimental to batteries. Batteries should always be stored away from
sources of heat whether in the unit or outside. The batteries contain a safety vent
which is designed to release gasses if the temperature of the batteries exceeds 125
degrees Fahrenheit. If the vent has been actuated the batteries should be replaced.
7. With proper care, the batteries used by the AutoCAT
provide many years of trouble free service. It is not required that these batteries be
replaced based solely on the basis of their age. However, to insure the highest level
of reliability, Arrow International recommends that the batteries be replaced after
three years of service. As the batteries age, it is important that their performance be
periodically checked to insure that they will perform as intended. It is recom-
mended that a load test, as outlined in this manual, be performed by qualified
service personnel at twelve month intervals to ensure battery capacity and usability.
Any time that the batteries do not pass the load test they must be replaced.
If it is not possible to perform the load test, or if the batteries' capacity and
usability cannot be verified for any reason, the batteries must be replaced after
three years of use.
8. If the optional second battery is installed and replacement of the batteries are
required for any reason, always replace the batteries in pairs and with the proper
type and rating of battery and follow battery replacement procedure outlined in this
manual. Never replace only one battery.
9. Sealed lead acid batteries are manufactured of highly recyclable materials and
should be recycled whenever possible. Never dispose of batteries in fire. Doing so
could result in possible rupture or explosion of battery. For lead acid battery
disposition instructions contact your local Government Authorities or Arrow
International local representative.
1 0 . M a i n t e n a n c e a n d S e r v i c e
1 0 . 2 : S e r v i c e M a i n t e n a n c e P r o c e d u r e s
®
2 Series IABP System should
Sealed Lead Acid Battery Maintenance
10-13