Preparation - Arrow AutoCAT 2 Series Operating Manual

Intra-aortic balloon pump iabp system
Table of Contents

Advertisement

Detailed Operations in Operator mode:

Preparation

The Operational Checkout procedure described in Section 10.1 is designed to verify the
AutoCAT
®
2 Series operational readiness. The instructions in this chapter assume that
the Operational Checkout has been performed recently (within a week). Follow the
preparation instructions below while the physician is inserting the balloon. Also inspect
all cables to insure proper operation.
If rapid initiation of counterpulsation is required, see "The AutoCAT
Emergency Conditions" on page 5.20 to set up the AutoCAT
AutoPilot
mode and begin pumping as quickly as possible.
1. Power ON
• Press the power switch located in the center of the patient interface panel. The power
switch LED should illuminate.
The AutoCAT
®
2 Series should already be plugged into a grounded AC power supply.
If the system is not secured, engage the wheel brakes with your feet to secure the
AutoCAT
®
2 Series.
• If the "SYSTEM RUNNING ON BATTERY" message is displayed, the AutoCAT
Series is not receiving AC power. Call a qualified person to check the AC outlets for
power. A fully charged battery will power the AutoCAT
of 90 minutes.
NOTE: Battery light "on" indicates battery at 80% or greater when connected to AC power.
• Double-check the helium and recorder paper supplies. These should have been
replenished during the Operational Checkout.
If there is 100psi or less of helium a low Helium tank pressure message will appear.
Refer to the instructions in Section 10.1 to replace the helium tank.
Be prepared to maintain IABP operation in critical situations by having a backup IABP system and extra
helium bottles ready in case of system failure or helium depletion.
Be prepared to change Operations modes if the currently selected Operation mode does not provide
Do not use solvents (e.g., acetone or other degreasing agents) to prepare the skin. They may damage the IAB
WARNING
WARNING
adequate assist.
WARNING
catheter or other plastic components of the system.
6 . O p e r a t i n g P r o c e d u r e s
6 . 1 : O p e r a t i n g I n s t r u c t i o n s
®
2 Series in
2 Series using the
®
®
®
2 Series for a minimum

Preparation

2
6-5

Advertisement

Table of Contents
loading

Related Products for Arrow AutoCAT 2 Series

This manual is also suitable for:

Autocat 2Autocat 2 wave

Table of Contents