Teledyne API Model 200A NO
Table 5-8: Status Output Pin Assignments
Output #
1
2
3
4
5
6
7
8
9
10
11
12
The Status Board schematic can be found in the Appendix.
5.5 RS-232 Interface
The RS-232 communications protocol allows the instrument to be connected to a wide variety of
computer based equipment. The interface provides two basic functions in the M200A.
1.
First is a comprehensive command interface for operating and diagnosing the analyzer in
interactive fashion. This mode is for use by a human operator issuing commands to the
instrument.
2.
Second is the equivalent command interface described above, but with the interface set
up to be operated by a computer program rather than interactive commands.
3.
The interface can also provide an audit trail of analyzer events. In this function the port
sends out messages about instrument events like calibration or warning messages. If these
messages are captured on a printer or remote computer, they provide a continuous audit trail
of the analyzers operation and status.
Analyzer Instruction Manual, 02246, Rev. F1
X
Pin #
Definition
1,2
ZERO CAL
3,4
SPAN CAL
5,6
FLOW ALARM
7,8
TEMP ALARM
9,10
DIAG MODE
11,12
POWER OK
13,14
SPARE
15,16
SPARE
17,18
SPARE
19,20
AUTORANGE - HI
21,22
SYSTEM OK
23,24
RX CELL PRESS
Condition
CLOSED IN ZERO CAL
CLOSED IN SPAN CAL
CLOSED IF FLOW WARNING
CLOSED IF ANY TEMP WARNING
CLOSED IN DIAG MODE
CLOSED IF SYSTEM POWER OK
CLOSED IF IN HIGH RANGE
CLOSED IF NO FAULTS PRESENT
CLOSED IF ABS PRES > 15" HG