5.9 System Test

Setting
Description
Blower
On = Blower on.
Off = Blower off.
V BattMeasEn
On = Sensor Readings screen displays battery voltage.
Off = Sensor Readings screen displays battery temperature.
V Out Enable
Reserved for future use.
PWRSWCLK
On = Clocks "1" to the STANDBY switch latch.
Off = No change.
PWRTUIOFF
On = Enters STANDBY mode (removes 18V_Boost, shutting down power).
Off = No change.
REF 1029568 Rev B
The System Test screen (Figure 5-15) includes:
Alarms: toggle the primary alarm speaker, backup alarm speaker, and
remote alarm relay ON or OFF.
Indicators: toggle the Alarm LED, Silence LED, and On Battery LED
ON or OFF.
Miscellaneous: toggle miscellaneous settings ON or OFF (Table 5-2).
When V BattMeasEn is On, the Sensor Readings screen displays
battery voltage; when Off, the Sensor Readings screen displays the
internal battery thermistor temperature.
Figure 5-15: Diagnostic Mode: System Test Screen
Table 5-2: System Test Screen: Miscellaneous Settings
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Diagnostic Mode and Troubleshooting
Chapter 5
5-15