Event ID
82
IPC COMMS FAILURE
83
IPC CKSUM FAILURE
84
TUI SOFT RESET
85
MODE SETTING FAIL
86
ST IPAP SETTING FAIL
87
ST EPAP SETTING FAIL
88
ST RISE SETTING FAIL
89
ST RATE SETTING FAIL
90
ST I-TIME SETTING FAIL
91
ST RAMP TIME SETTING FAIL Controller processor detected that the ST mode ramp time setting change is not
REF 1029568 Rev B
Table 5-3: Significant Event Codes (Continued)
Comment / Corrective Action
Controller processor detected an IPC message failure.
1. Check the interprocessor communications cable.
2. If the problem persists replace the controller PCB.
3. Retrieve the DRPT and contact technical support.
Controller processor detected an IPC message with an invalid checksum.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an IPC soft reset message which was sent from the
user interface processor.
1. If the problem persists replace the controller PCB.
2. Replace the user interface PCB.
3. Retrieve the DRPT and contact technical support.
Controller processor detected that the mode setting change is not consistent with
the value stored in EEPROM.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected that the ST mode IPAP setting change is not
consistent with the value stored in EEPROM.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected that the ST mode EPAP setting change is not
consistent with the value stored in EEPROM.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected that the ST mode rise time setting change is not
consistent with the value stored in EEPROM.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected that the ST mode breath rate setting change is not
consistent with the value stored in EEPROM.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected that the ST mode inspiratory time setting change is
not consistent with the value stored in EEPROM.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
consistent with the value stored in EEPROM.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Diagnostic Mode and Troubleshooting
Chapter 5
5-25