Real-Time Clock Calibration; Downloading Real-Time Clock Calibration Software - Respironics BiPAP Service Manual

Table of Contents

Advertisement

Chapter 7
Testing, Calibration, and Performance Verification
7.5 Real-time Clock
Calibration
7-6
Calibration software compares the ventilator real-time clock to service PC time
to determine exactly when an event occurs, Because the calibration fails if the
times do not match, perform the real-time clock calibration before you proceed
to the calibration procedure.
7.5.1

Downloading Real-Time Clock Calibration Software

Follow these steps to download the most current software to the service PC,
then perform the real-time clock calibration:
1. From the service PC, go to http://my.respironics.com and click on
Service Software (Figure 7-1).
Figure 7-1: Select Service Software
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B