REF 1029568 Rev B
16. When software installation is complete click Next (Figure 6-38).
Figure 6-38: Software Installation Complete
17. When BiPAP Focus Wizard Phase 2 Completion window appears, click
Finish (Figure 6-39).
Figure 6-39: BiPAP Focus Wizard Phase 2 Completion window
18. The ventilator restarts in Normal mode using the default settings.
Depending on the ventilator configuration, you may need to perform a
calibration as described in Chapter 7.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Chapter 6
Ventilator Communications
6-21