Chapter 5
Diagnostic Mode and Troubleshooting
5-14
905
Pwr On reset
300
POST Successful
102
Power Off
639
Disconnect
905
Pwr On reset
300
POST Successful
9. Once HyperTerminal displays the DRPT in the main window, click
Transfer > Capture Text > Stop to finish capturing text to the text file.
NOTE:
The DRPT capture file is a TXT file by default: use any word processing
program to view the file.
Figure 5-14 shows how to interpret each line of the DRPT report.
437 Battery Voltage Failed
Battery Voltage Failed - description
437 - diagnostic Code
Figure 5-14: Interpreting the DRPT Report
NOTE:
The last two fields (if applicable) can also show mode/setting changes
and their from/to values.
10. To save the DRPT, highlight the appropriate fields (or Edit>Select All),
copy (Edit>Copy) and then paste into any text editor/word processing/
email program.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
6/13/09
16:32
6/13/09
16:32
6/13/09
16:32
6/13/09
16:31
6/13/09
16:27
6/13/09
16:27
6/19/09
20:42
V 615,615
V or T - System errors
V= voltage, T= temperature
(if applicable)
20:42 - Time stamp of occurrence
6/19/09 - MM/DD/YY
615, 615 - previous
and current values
REF 1029568 Rev B