Chapter 7
Testing, Calibration, and Performance Verification
5. Select SET RT CLOCK (Figure 7-5).
Figure 7-5: Selecting Set RT Clock
6. At the Real-Time Calibration Tool window, click the RT Clock to begin
calibration (Figure 7-6).
Figure 7-6: Real Time Clock Calibration Tool
7-9
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.