Ventilator Run-In - Respironics BiPAP Service Manual

Table of Contents

Advertisement

Chapter 7
Testing, Calibration, and Performance Verification

7.3 Ventilator Run-in

7-4
Perform the ventilator run-in procedure for at least 120 minutes to avoid
possible calibration failures.
Required equipment:
Test orifice, BiPAP, 0.25-in. (P/N 332353)
Follow these steps:
1. Connect the test orifice to the ventilator outlet port.
2. Run the ventilator for at least 2 hours (120 minutes) before
performing the calibration at default settings:
S/T mode
IPAP = 12 cmH
EPAP = 4 cmH
Rise-Time = 2
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
RR = 4/min
O
I-Time = 1 sec
2
O
Ramp Time = 0 min
2
Ramp Start = 4 cmH
O
2
REF 1029568 Rev B