Chapter 7
Testing, Calibration, and Performance Verification
7-14
4. At the Diagnostic mode main menu, highlight Product Information and
press Enter.
5. At the Product Information screen (Figure 7-11), verify the serial
number. If the serial number does not match, reprogram the serial
number as described in section 6.4.
Figure 7-11: Product Information Screen
6. Ensure the ventilator is in Diagnostic mode. Remove the bacteria filter
from the ventilator if connected.
Pressure calibration:
7. Connect the oxygen enrichment adapter to the patient outlet of the
ventilator, and use the silicone cork to occlude the adapter.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B