REF 1029568 Rev B
Testing, Calibration, and Performance Verification
8. Connect the pressure tubing from the oxygen port of the oxygen
enrichment adapter to the pneumatic calibration analyzer
(Figure 7-12).
Figure 7-12: Pressure Calibration Setup
9. On the service PC, open the Service Calibration Software and start the
calibration by clicking on the white Run arrow in the upper left corner
of the window (Figure 7-13).
Run
arrow
Figure 7-13: Starting Calibration
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Chapter 7
7-15