Respironics BiPAP Service Manual page 66

Table of Contents

Advertisement

Chapter 5
Diagnostic Mode and Troubleshooting
Event ID
636
ABNORMAL RESTART
637
BATT CHARGE FAILURE
638
INDICATOR FAILURE
639
DISCONNECT
640
HI P REG
641
LO P REG
642
LOSS OF AC
643
LOW BATTERY
644
APNEA
645
APNEA RATE
646
KEY STUCK
647
RTC FAILURE
648
HIGH TEMPERATURE
649
LOW TEMPERATURE
7XX: Unused codes.
8XX: User interface processor software anomaly codes.
801
Soft. Anomaly
802
TUI RTOS Error
803
Alarm Corrupt Stat.
804
Task Monitor
805
Unused
806
68332M Trap
807
TUI IPC Message too big
5-34
Table 5-3: Significant Event Codes (Continued)
Comment / Corrective Action
1. Retrieve the DRPT and contact technical support.
1. See alarms information in Operator's Manual.
1. See pre-operational check information in Operator's Manual.
1. See alarms information in Operator's Manual.
1. See alarms section of Operator's Manual.
1. See alarms information in Operator's Manual.
1. See alarms information in Operator's Manual.
1. See alarms information in Operator's Manual.
1. See alarms information in Operator's Manual.
1. See alarms information in Operator's Manual.
1. See alarms information in Operator's Manual.
1. Retrieve the DRPT and contact technical support.
1. See alarms information in Operator's Manual.
1. See alarms information in Operator's Manual.
Software anomaly error code that identifies a file and line number of the
assertion.
1. Retrieve the DRPT and contact technical support.
User interface RTOS error.
1. Retrieve the DRPT and contact technical support.
User interface processor detected an invalid/inconsistent alarm status.
1. Retrieve the DRPT and contact technical support.
The task monitor on the user interface processor detected a task that did not
check in within the scheduled time.
1. Retrieve the DRPT and contact technical support.
User interface processor encountered a processor trap.
1. Retrieve the DRPT and contact technical support.
An inter-process communication (IPC) message sent between the user interface
and controller processors is too large.
1. Retrieve the DRPT and contact technical support.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B

Advertisement