Respironics BiPAP Service Manual page 112

Table of Contents

Advertisement

Chapter 7
Testing, Calibration, and Performance Verification
7-18
Positive flow calibration:
19. The window (Figure 7-17), indicates the start of the flow calibration,
beginning with positive flow.
20. Press ENTER on the keyboard. Unblock the oxygen enrichment
adapter, then connect patient tubing or a coupling between the oxygen
enrichment adapter and the flowmeter inlet. Connect another length of
patient tubing from the flowmeter outlet of the to the control valve
(completely open the flow valve) (Figure 7-18).
Figure 7-18: Positive Flow Calibration Measurement Setup
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Figure 7-17: Flow Calibration
REF 1029568 Rev B