Extended Self Test
(EST)
REF 580-1000-01 Y
Press the OK button to clear the message from the screen. You may continue
to use the ventilator without the backup battery or power off the ventilator and
reconnect the backup battery.
WARNING:
Always turn the ventilator power OFF before connecting additional
equipment.
To turn this feature off, enter Diagnostic Mode and deactivate the messaging
feature by pressing the Bkup Battery button.
1. Power off the ventilator.
2. Power on the ventilator while holding down the ALARM RESET and
100% O
keys for approximately 5 seconds.
2
3. A message appears on the ventilator screen asking the user to 'Verify
that the patient is disconnected prior to proceeding.' Press OK to enter
Diagnostic Mode.
4. Once you've entered Diagnostic Mode, select USER CONFIG button
which takes you to the User Config screen.
5. Press the BKUP BATTERY button. The background color should return
to blue signifying that this feature has been deactivated.
Respironics recommends that the user runs Extended Self Test (EST) upon
receipt of the ventilator to ensure that there has been no shipping damage to
the system. Respironics also recommends that the user run EST between
patients to verify the overall functional integrity of the ventilator. Refer to
chapter 11 for instructions on running EST.
If EST was run successfully and all configuration information has been
entered, Esprit is ready to be used. Follow the recommended "Preoperational
Procedure" on page 8-30 once a patient has been selected.
To exit the Diagnostics Mode you must turn the Power On/Off switch OFF (
then ON (
).
Esprit® Ventilator Operator's Manual © Respironics, Inc.
Chapter 5
Setup
)
5-13