Respironics BiPAP Service Manual page 110

Table of Contents

Advertisement

Chapter 7
Testing, Calibration, and Performance Verification
7-16
10. At the prompt, enter the ventilator serial number exactly as it appears,
then click OK.
11. At the prompt, enter the ventilator model number and click OK.
12. Enter operator name and click OK.
13. Verify that the label is BiPAP Focus and click PASS.
14. The program runs an alarm test: follow the onscreen instructions.
15. At the prompt (Figure 7-14), attach the manometer and occlude the
test orifice, then click OK to start sensor calibration.
Figure 7-14: Pressure Sensor Calibration
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B