Battery Replacement - Respironics BiPAP Service Manual

Table of Contents

Advertisement

REF 1029568 Rev B
4.4.3

Battery Replacement

Follow these steps to replace the battery (Figure 4-2):
1. Turn the system OFF (O) and disconnect the DC power supply.
2. Remove the two screws that secure the battery compartment door.
3. Remove the old battery: remove from the compartment and unplug the
cable.
4. Install the new battery: plug in the cable and install the battery into
the compartment.
5. Replace the battery compartment door and screws.
6. Reconnect the DC power supply.
CAUTION:
Use only Respironics-approved batteries.
7. Allow sufficient time to fully charge the new battery (Battery is fully
charged when the Battery Charging LED turns off).
Battery compartment
Battery cable
connector
Battery
compartme
nt door
Figure 4-2: Replacing the Battery
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Chapter 4
Periodic Maintenance
4-5

Advertisement

Table of Contents
loading

Table of Contents