Specifications; Ventilator Specifications - Respironics BiPAP Service Manual

Table of Contents

Advertisement

Chapter 10. Specifications
10.1 Ventilator

Specifications

REF 1029568 Rev B
Default Settings
S/T mode
RR = 4/min
IPAP = 12 cmH
O
I-Time = 1 sec
2
EPAP = 4 cmH
O
Ramp Time = 0 min
2
Rise-Time = 2
Ramp Start = 4 cmH
Settings: CPAP Mode
CPAP
Range: 4-20 cmH
Resolution: 1 cmH
Dynamic accuracy: ± 5 cmH
Ramp Time
Range: 0-45 min.
Resolution: 5 min.
Accuracy: ±10% of setting
Ramp Start
Range: 4 cmH
Resolution: 1 cmH
Settings: S/T Mode
IPAP
Range: 4-30 cmH
Resolution: 1 cmH
Dynamic accuracy: ± 5 cmH
IPAP cannot be set below EPAP.
EPAP
Range: 4-25 cmH
Resolution: 1 cmH
Dynamic accuracy: ± 5 cmH
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
O
2
O (4-20 hPa)
2
O (1 hPa)
2
O (5 hPa)
2
O to CPAP setting (4 hPa to CPAP setting)
2
O (1 hPa)
2
O (4-30 hPa)
2
O (1 hPa)
2
O (5 hPa)
2
O (4-25 hPa)
2
O (1 hPa)
2
O (5 hPa)
2
10-1