Event ID
1
GENERIC SOFTWARE ERROR
2
SOFTWARE CORRUPT
3
EXTERNAL RAM FAILURE
4
FIQ STACK OVERFLOW
5
FIQ STACK UNDERFLOW
6
NESTED IRQ STACK
OVERFLOW
7
NESTED IRQ STACK
UNDERFLOW
8
IRQ STACK OVERFLOW
REF 1029568 Rev B
Use the arrow keys to scroll through the log.
corrective actions.
When performing corrective actions:
NOTE:
Perform the corrective actions in the order listed.
Always retrieve and save the DRPT if possible before replacing
parts. If the corrective action suggests swapping the user
interface or controller PCB, retrieve the DRPT before installing
the replacement PCB.
To verify that replacing a part corrects the problem,
temporarily install (swap) a replacement part. If the
replacement corrects the problem, verify that the problem
recurs with the original part before replacing it.
Table 5-3: Significant Event Codes
Comment / Corrective Action
Controller processor detected a software error.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a program memory checksum error.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an external RAM memory error.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a stack overflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a stack underflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a stack overflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a stack underflow.
1. Retrieve the DRPT and contact technical support.
Controller processor detected a stack overflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Diagnostic Mode and Troubleshooting
lists event IDs and
Table 5-3
Chapter 5
5-17