Emissions Test/Compliance
RF emissions CISPR 11/Class B
Harmonic emissions IEC 61000-3-2/
(not applicable)
Voltage fluctuations/flicker emissions
IEC 61000-3-3/ (not applicable)
Guidance and manufacturer's declaration: electromagnetic immunity, IEC
60601-1-2 + A1:2000.
The BiPAP Focus System is suitable for use in the specified electromagnetic
environment. The customer and/or the user of the BiPAP Focus System should
ensure that it is used in an electromagnetic environment as described below:
IEC 60601
Immunity Test
Test Level
Electrostatic
± 6 kV
discharge (ESD)
contact
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for
transient/burst
power supply
IEC 61000-4-4
lines
± 1 kV for
input/output
lines
Surge
± 1 kV
IEC 61000-4-5
differential
mode
± 2 kV
common
mode
NOTE: UT is the AC mains voltage prior to application of the test level.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Electromagnetic Environment
Guidance
The BiPAP Focus System is suitable
for use in all establishments other
than domestic and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Compliance
Electromagnetic Environment
Level
Guidance
± 6 kV
Floors should be wood, concrete, or
contact
ceramic tile. If floors are covered
with synthetic material, the relative
± 8 kV air
humidity should be at least 30%.
± 2 kV for
Mains power quality should be that
power supply
of a typical commercial and/or
lines
hospital environment.
± 1 kV for
input/output
lines
± 1 kV
Mains power quality should be that
differential
of a typical commercial and/or
mode
hospital environment.
± 2 kV
common
mode