10.7 Warranty

REF 1029568 Rev B
Recommended separation distances between portable and mobile RF communications
equipment and the BiPAP Focus IEC 60601-1-2
Frequency of
150 kHz to
transmitter
80 MHz
Equation
d = 1.2
Rated maximum
Separation
output power of
distance (m)
transmitter (W)
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the
separation distance can be estimated using the equation in the corresponding
column, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects, and
people.
Respironics warrants the BiPAP Focus Ventilator to be free from defects in
material and workmanship for a period of 12 months from the date of
purchase, provided that the unit is operated under conditions of normal use as
described in this manual. At its discretion, Respironics will make
replacements, repairs, or issue credits for equipment or parts that are found to
be defective.
THE WARRANTY SET FORTH ABOVE IS THE SOLE AND EXCLUSIVE
WARRANTY WITH RESPECT TO THE PRODUCT AND RESPIRONICS DOES
NOT MAKE, AND HEREBY SPECIFICALLY DISCLAIMS, ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION,
ANY IMPLIED WARRANTY OF MERCHANTABILITY OR IMPLIED WARRANTY
OF FITNESS FOR A PARTICULAR PURPOSE. THE REPAIR, REPLACEMENT
OR CREDIT REMEDY SET FORTH ABOVE WILL BE THE SOLE REMEDY FOR
BREACH OF WARRANTY. IN NO EVENT SHALL RESPIRONICS BE LIABLE
FOR LOST PROFITS, LOSS OF GOOD WILL, OR INCIDENTAL OR
CONSEQUENTIAL DAMAGES EVEN IF RESPIRONICS HAS BEEN ADVISED
OF THE POSSIBILITY OF THE SAME.
Exclusions
This warranty does not apply to any unit or individual parts which have been
repaired or altered in any way that in Respironics' judgement, affect its ability
or reliability, or which has been subjected to misuse, negligence, abuse, or
accident.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
:
150 kHz to
800 MHz
P
P
d = 1.2
Separation
distance (m)
0.12
0.38
1.2
3.8
12
Chapter 10
Specifications
800 MHz to 2.5
GHz
P
d = 2.3
Separation
distance (m)
0.23
0.73
2.3
7.3
23
10-13