Respironics BiPAP Service Manual page 50

Table of Contents

Advertisement

Chapter 5
Diagnostic Mode and Troubleshooting
Event ID
9
IRQ STACK UNDERFLOW
10
TIMER STACK OVERFLOW
11
TIMER STACK UNDERFLOW
12
SERVICE STACK OVERFLOW
13
SERVICE STACK UNDERFLOW Controller processor detected a stack underflow.
14
THREAD STACK OVERFLOW
15
UNDEFINED INSTRUCTION
16
UNEXPECTED SOFTWARE
INTERRUPT
17
MICROPROCESSOR
PREFETCH EXCEPTION
18
DATA ACCESS EXCEPTION
19
RESERVED EXCEPTION
20
SPURIOUS DEFAULT
INTERRUPT
21
SPURIOUS INTERRUPT
5-18
Table 5-3: Significant Event Codes (Continued)
Comment / Corrective Action
Controller processor detected a stack underflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a stack overflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a stack underflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a stack overflow.
1. Retrieve the DRPT and contact technical support.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected a stack overflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an unexpected exception.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an unexpected exception.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an unexpected exception.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an unexpected exception.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an unexpected exception.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an unexpected interrupt.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an unexpected interrupt.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B