Chapter 5
Diagnostic Mode and Troubleshooting
Event ID
31
UN-REPAIRABLE
PARAMETER STORAGE
32
CORRUPT PARAMETER
STORAGE
33
TCB NVDATA STORAGE
ERROR
34
UNABLE TO QUEUE DATA TO
PARAMETER STORAGE
35
UNABLE TO WRITE TO
PARAMETER STORAGE
36
PARAMETER OUT OF RANGE
IN PARAMETER STORAGE
37
CORRUPT REAL-TIME CLOCK
VALUE
38
REAL-TIME CLOCK NOT
TICKING
39
UNABLE TO QUEUE DATA TO
USER INTERFACE
40
INVALID BUILT-IN SELF TEST
CALL
5-20
Table 5-3: Significant Event Codes (Continued)
Comment / Corrective Action
Controller processor detected an error accessing the calibration data in
EEPROM.
1. Attempt to recalibrate the system.
2. If calibration fails, replace the controller PCB.
3. Retrieve the DRPT and contact technical support.
Controller processor detected a checksum error on the calibration in EEPROM.
1. Attempt to recalibrate the system.
2. If calibration fails, replace the controller PCB.
3. Retrieve the DRPT and contact technical support.
Controller processor detected that the data structure is bigger than the EEPROM
device.
1. Attempt to recalibrate the system.
2. If calibration fails, replace the controller PCB.
3. Retrieve the DRPT and contact technical support.
Controller processor detected that the queue to the EEPROM manager thread is
full.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected that a write operation to EEPROM failed the
readback check.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected that a write operation to EEPROM failed the range
check for that data item.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected that the real-time clock (RTC) time and the RTOS
tick timer have drifted since it was last tested. When changing the time, this
could happen and an RTC Failure alarm may be annuciated. This alarm can be
reset by pressing Alarm Reset. No action required.
Controller processor detected that the RTC time and the RTOS tick timer have
drifted significantly (>4 seconds) since it was last tested. When changing the
time, this could happen and an RTC Failure alarm may be annuciated. This
alarm can be reset by pressing Alarm Reset.
1. Retrieve the DRPT and contact technical support.
Controller processor detected a RTOS queue full condition.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Controller processor detected an invalid BIST test was attempted to be run.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
BiPAP Focus Ventilator Service Manual © Respironics, Inc.