Table of Contents

Advertisement

Quick Links

BiPAP Focus Ventilator
S E R V I C E M A N U A L

Advertisement

Table of Contents
loading

Summary of Contents for Respironics BiPAP

  • Page 1 BiPAP Focus Ventilator S E R V I C E M A N U A L...
  • Page 2 ALL RIGHTS RESERVED. This work is protected under Title 17 of the United States copyright code and is the sole property of Respironics. No part of this document may be copied or otherwise reproduced, or stored in any electronic information retrieval system, except as specifically permitted under United States copyright law, without the prior written consent of Respironics.
  • Page 3: Table Of Contents

    Storing the Ventilator ........4-9 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 4 Installing Operating Software on the Ventilator ....6-12 Reprogramming the BiPAP Focus Serial Number....6-22 7 Testing, Calibration, and Performance Verification .
  • Page 5 8.11 Interconnect Diagrams ........8-18 9 BiPAP Focus Replacement Parts List ......9-1 Complete Parts List .
  • Page 6 Table of Contents BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 7: Introduction And Intended Use

    The BiPAP Focus is appropriate for acute, sub-acute, and intra-hospital transport settings. The BiPAP Focus is intended for use with nasal and full face masks, and is not intended to be a life-support device.
  • Page 8: Tools And Test Equipment

    Hand Tools and Materials Antistatic workstation 3M Model 725 or equivalent Cable kit, BiPAP Focus, communications (USB and Respironics P/N 1039800 serial cables) Cable, BiPAP Focus, battery charging (NiMH to 9V Respironics P/N 1040421 alkaline) Cleaning cloth Local supplier Isopropyl alcohol...
  • Page 9 1.2 Where to Go for For customer service and product support contact: Help Respironics, Inc. USA telephone: 1-800-345-6443 Fax: 1-800-886-0245 International telephone: 724-387-4000 International fax: 724-387-5012 service@respironics.com. clinical@respironics.com. www.respironics.com. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 10 Chapter 1 Introduction and Intended Use (This page is intentionally blank.) BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 11: Warnings And Cautions

    • To avoid injury to the patient airway, DO NOT use the BiPAP Focus where ambient room temperature is above 95 °F (35 °C) because outlet gas temperature may exceed 106° F (41°C).
  • Page 12 Explosion hazard. Do not operate the ventilator in the presence of flammable anesthetic agents. • To avoid the risk of fire, use only Respironics-approved batteries. All battery connectors are keyed to ensure proper connection. Do not attempt to connect a battery incorrectly.
  • Page 13: Cautions

    If a 343 error code is displayed, remove the battery and charge it using the BiPAP Focus battery charging cable (P/N 1040421). This cable provides the initial charge required so that the ventilator can recharge a depleted battery. Refer to the battery charge instructions supplied with the cable.
  • Page 14: Notes

    Chapter 2 Warnings and Cautions 2.3 Notes • Refer to the BiPAP Focus Operator’s Manual for product use, additional warnings, cautions, and notes. • Service calibration procedures require a printer to be assigned to the service PC: this can be an internal print location (for example, using Adobe PDF) or a printer connected to the PC.
  • Page 15: Theory Of Operation

    AC power conditions occur. The system displays a real- time estimated delivered pressure bar graph. The BiPAP Focus includes alarm silence and alarm pre-silence features. A fully charged backup battery provides a minimum of 45 minutes of backup power at default settings in case AC power is not available (for example, during transport within the hospital).
  • Page 16: Pneumatic System

    Chapter 3 Theory of Operation 3.1 Pneumatic Figure 3-1 shows a pneumatic diagram of the BiPAP Focus ventilator. System Diffuser BiPAP Focus ventilator Bacteria filter Air inlet (mandatory) Flow sensor filter(s) Patient circuit (to mask outlet) Blower Flow Muffler (motor)
  • Page 17: Blower

    3.1.6 Diffuser The diffuser is located on the output of the flow valve and is designed to provide a reduction of air turbulence. Diffuser REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 18: Electronic System

    Chapter 3 Theory of Operation 3.2 Electronic System The BiPAP Focus electronic system controls, monitors, and displays patient and ventilator data. Two microcontrollers (on the user interface and controller PCBs) interpret user and sensor inputs while controlling inspiratory and expiratory pressures according to the selected ventilator settings. Figure 3-2 shows an interconnect block diagram of the BiPAP Focus ventilator.
  • Page 19: User Interface Pcb

    J10 of the controller PCB 18 VDC to controller PCB ON/OFF switch serial power supply port connection port Figure 3-3: BiPAP Focus User Interface PCB REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 20: Controller Pcb

    Flow valve assembly J9 Interprocessor Interprocessor communications communications cable J10 cable J1 Connects to user Connects to user interface PCB (J1) interface PCB (J1) Figure 3-4: BiPAP Focus Controller PCB BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 21: Vga Interface Pcb

    9.6 V, 3.7 Ah, nickel metal hydride (NiMH) • In-line thermal fuse • Built-in thermistor for temperature monitoring • Nominal charge time under 5 hours • Nominal run time 45 minutes at default settings REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 22: Back Panel

    Power cord strain relief clip Figure 3-5: Back Panel NOTE: The ON/OFF switch controls all power management function. Internal battery charging and ventilator operation are disabled in the OFF (O) position. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 23: Periodic Maintenance

    Chapter 4. Periodic Maintenance This chapter describes cleaning and maintenance procedures for the BiPAP Focus ventilator. To ensure correct operation, perform maintenance at the recommended intervals. Table 4-1 summarizes BiPAP Focus maintenance procedures. Table 4-1: BiPAP Focus Periodic Maintenance Part...
  • Page 24: Cleaning The Reusable Air Inlet Filter

    Replace the reusable air inlet filter every year or if damage is visible. Install only 1 reusable filter. CAUTION: Replace the disposable air inlet filter if debris or damage is visible. Do not attempt to clean or reuse. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 25: Battery Maintenance

    Chapter 4 Periodic Maintenance 4.4 Battery WARNING: To avoid the risk of fire, use only Respironics-approved batteries. All battery connectors are keyed to ensure proper connection. Do not attempt to connect Maintenance a battery incorrectly. NOTE: Recharge the battery if a Low Battery or Battery Depleted alarm occurs during a pre-operational check.
  • Page 26: Battery Charging

    Battery Depleted or Missing alarm. If a Battery Depleted or Missing alarm occurs, connect AC power immediately to avoid total loss of power. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 27: Battery Replacement

    7. Allow sufficient time to fully charge the new battery (Battery is fully charged when the Battery Charging LED turns off). Battery compartment Battery cable connector Battery compartme nt door Figure 4-2: Replacing the Battery REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 28: Fully Depleted Battery

    Because reconnecting the charged battery is time-sensitive, read these instructions before beginning this procedure. BiPAP Focus charging circuitry is designed to detect a battery with a minimum voltage of 2.80 VDC as measured across the red and black battery connector wires.
  • Page 29 Figure 4-4: Removing the Battery Compartment Screws 3. Remove the internal battery and disconnect its cable (Figure 4-5). Disconnect Figure 4-5: Removing the Internal Battery REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 30 Figure 4-6: Charging the Battery 5. Allow the two to remain connected for a minimum of 15 minutes. 6. Disconnect the battery charging cable and reinstall the BiPAP Focus battery within 30 seconds. 7. Connect the ventilator AC power cord to wall power and turn the ON/ OFF switch ON (I).
  • Page 31: Storing The Ventilator

    Store the ventilator in a location that meets environmental specifications. To avoid premature battery depletion, ensure that the ventilator power switch is in Ventilator the OFF (O) position. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 32 Chapter 4 Periodic Maintenance (This page is intentionally blank.) 4-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 33: Diagnostic Mode And Troubleshooting

    15 seconds or until Diagnostic mode appears on the screen. 2. At the warning prompt (Figure 5-1), press OK to start Diagnostic mode. The Diagnostic mode main menu (Figure 5-2) appears. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 34 3. Press the arrow keys to highlight the menu selection, then press Enter. 4. To exit Diagnostic mode at any time, turn the ventilator off using the STANDBY key. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 35: Product Information

    This serial number indicates a manufacture date of 29 July 2006: S/N 8772 060729-01 P/N 1028772 Unit number Date: YY/MM/DD Include the hyphen if present (varies by build date) when recording the serial number REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 36: System Settings

    Once a new language is selected, a window appears notifies you that the language is being updated, and that the ventilator will shut down in 5 seconds. The new language is displayed when the ventilator restarts. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 37: Preferences

    Use Reverse Video to toggle between reverse and normal video. • Use Screen Lock Timer to turn the screen lock on or off. 3. Press the Menu key to return to the main menu. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 38: Significant Event Log

    (section 5.10 describes significant event codes in detail). Figure 5-6: Diagnostic Mode: Significant Event Log Use the arrow keys to scroll up and down. Press Menu to return to the main menu. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 39: Sensor Readings

    +18V (mains input) measurement. Range: 882 to 1008 counts Voltage(V) = counts * 0.019043 +18V Boost +18V_Boost supply measurement. Range: 923 to 1020 counts. Voltage(V) = counts * 0.019043 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 40 Status of battery top-off charge. 1: Battery NOT in TOC state 0: Battery in TOC state BattChrgRdy Status of battery ready to charge. 1: Ready for charging 0: Not ready for charging BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 41: Setting Up The Serial Interface For Drpt

    3. Launch the HyperTerminal program on the PC by clicking Start > Programs > Accessories > Communications > HyperTerminal. 4. Enter a name for the connection (Figure 5-8 shows BiPAP Focus), choose an icon, then click OK. REF 1029568 Rev B...
  • Page 42 Chapter 5 Diagnostic Mode and Troubleshooting Figure 5-8: Entering a Name for the Connection to the BiPAP Focus Ventilator 5. Select the appropriate serial port from the Connect using: drop-down list (Figure 5-9 shows COM1), then click OK. Figure 5-9: Selecting the Serial Port 6.
  • Page 43 10. Click the Terminal Setup button. 11. Select underline and blink for cursor settings then click OK. Figure 5-12: Terminal Settings 12. Click the ASCII Setup button (Figure 5-11). 5-11 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 44 13. Match the ASCII Setup screen (Figure 5-13), then click OK. Figure 5-13: ASCII Setup Screen 14. Save this configuration. Select File > Save As and save to the desktop if you want to create an icon for BiPAP Focus Communications on the Windows desktop. 5-12 BiPAP Focus Ventilator Service Manual ©...
  • Page 45: Generating A Diagnostic Report (Drpt)

    Chapter 5 Diagnostic Mode and Troubleshooting 5.8 Generating a 1. With the BiPAP Focus ventilator in Diagnostic mode and the Diagnostics end of the serial communications cable connected to the Diagnostic Report ventilator, start HyperTerminal, open the BiPAP Focus (DRPT) Communications file, or click on the BiPAP Focus Communications icon on the PC (if created).
  • Page 46 10. To save the DRPT, highlight the appropriate fields (or Edit>Select All), copy (Edit>Copy) and then paste into any text editor/word processing/ email program. 5-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 47: System Test

    On = Clocks “1” to the STANDBY switch latch. Off = No change. PWRTUIOFF On = Enters STANDBY mode (removes 18V_Boost, shutting down power). Off = No change. 5-15 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 48: Significant Event Codes

    903 - Ext Wdog Unexpected and 310 POST Abnormal Reset.) Off = No change. 5.10 Significant The BiPAP Focus Ventilator generates a log of system events that can be used for troubleshooting, including settings changes, alarms, and codes. If a Event Codes...
  • Page 49 1. Retrieve the DRPT and contact technical support. IRQ STACK OVERFLOW Controller processor detected a stack overflow. 1. If the problem persists replace the controller PCB. 2. Retrieve the DRPT and contact technical support. 5-17 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 50 2. Retrieve the DRPT and contact technical support. SPURIOUS INTERRUPT Controller processor detected an unexpected interrupt. 1. If the problem persists replace the controller PCB. 2. Retrieve the DRPT and contact technical support. 5-18 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 51 Controller processor detected an error in format of the calibration data in PARAMETER STORAGE EEPROM. 1. Attempt to recalibrate the system. 2. If calibration fails, replace the controller PCB. 3. Retrieve the DRPT and contact technical support. 5-19 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 52 Controller processor detected an invalid BIST test was attempted to be run. CALL 1. If the problem persists replace the controller PCB. 2. Retrieve the DRPT and contact technical support. 5-20 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 53 1. Replace the controller PCB. 2. If the problem persists replace the pneumatics subassembly (blower and valve). 3. Retrieve the DRPT and contact technical support. 5-21 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 54 UNABLE TO QUEUE DATA TO Controller processor detected a RTOS queue full condition. PROVIDE THERAPY 1. Replace the controller PCB. 2. Retrieve the DRPT and contact technical support. 5-22 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 55 OPERATING SYSTEM FAILED Controller processor detected an initialization failure when starting the RTOS. INITIALIZATION 1. Replace the controller PCB. 2. Retrieve the DRPT and contact technical support. 5-23 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 56 INVALID IPC MESSAGE ID Controller processor detected an IPC message with an invalid message id. 1. Replace the controller PCB. 2. Retrieve the DRPT and contact technical support. 5-24 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 57 ST RAMP TIME SETTING FAIL Controller processor detected that the ST mode ramp time setting change is not consistent with the value stored in EEPROM. 1. Replace the controller PCB. 2. Retrieve the DRPT and contact technical support. 5-25 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 58 2XX: boot block error codes. Not logged to the significant event log because they are detected very early in the power on sequence. The system annunciates a system error alarm and displays the error code and the text on the screen. 5-26 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 59 POST TCB NOT RESPONDING During POST the user interface processor detected that the controller processor timed out. 1. Check the interprocessor communications cable. 2. If the problem persists replace the controller PCB. 3. Replace the user interface PCB. 304-307 Unused 5-27 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 60 12V VOLTAGE RAIL FAILED 3. Enter Diagnostics mode and use the Sensor Readings screen to check the voltage and ADC readings. 4. Replace the user interface PCB. 5-28 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 61 BOARD TEMPERATURE TOO specifications. 2. Enter Diagnostics mode and use the Sensor Readings screen to check the voltage and ADC readings. 3. Replace the user interface PCB. 5-29 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 62 1. Ensure that the system is at room temperature or within operating specifications. 2. Enter Diagnostics mode and use the Sensor Readings screen to check the temperature and ADC readings. 3. Replace the user interface PCB. Unused 5-30 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 63 2. Check if the problem occurs during AC and battery operation. 3. Enter Diagnostics mode and use the Sensor Readings screen to check the voltage and ADC readings. 4. Replace the user interface PCB. 5-31 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 64 1. Ensure that the system is at room temperature or within operating specifications. 2. Enter Diagnostics mode and use the Sensor Readings screen to check temperature and ADC readings. 3. Replace the user interface PCB. 5-32 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 65 No action required FROM TIME AND DATE CHANGE TO No action required 6XX: Ventilator alarm codes. 600-634 Unused BATTERY DEPLETED 1. See alarms information in Operator's Manual. 5-33 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 66 TUI IPC Message too big An inter-process communication (IPC) message sent between the user interface and controller processors is too large. 1. Retrieve the DRPT and contact technical support. 5-34 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 67 TCB SOFT RESET Controller PCB has detected a software reset condition and attempted to send it to the user interface processor. 1. Retrieve the DRPT and contact technical support. 5-35 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 68 Chapter 5 Diagnostic Mode and Troubleshooting (This page is intentionally blank.) 5-36 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 69: Ventilator Communications

    Chapter 6. Ventilator Communications This chapter describes how to install operating software onto the BiPAP Focus ventilator. 1. Connect the service PC to the ventilator and enter ventilator Download mode (section 6.1). 2. Install USB driver software onto the service PC (section 6.2).
  • Page 70: Connecting The Service Pc And Ventilator

    NOTE: Do not connect or use the BiPAP Focus USB cable while connected to a docking station. Do not connect the BiPAP Focus USB cable through a USB hub. Using docking stations or USB hubs causes communication errors.
  • Page 71 Chapter 6 Ventilator Communications 2. Connect the USB cable (from BiPAP Focus communications cable kit P/N 1039800): insert the Type B end into the ventilator USB port, and the Type A end to the service PC (Figure 6-2). USB Type B end...
  • Page 72 Alarm sounds and WAITING FOR DOWNLOAD SERVER appears 1. Press the STANDBY key Figure 6-4: Entering Download Mode 5. Install USB driver software as described in section 6.2. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 73: Loading Usb Drivers To The Service Pc

    3. At Can Windows connect to Windows Update to search for software? click Yes, this time only then Next. 4. Select Install the Software Automatically (Recommended) then click Next (Figure 6-6). The Wizard begins searching. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 74 5. At the prompt indicating that the software you will be installing has not passed Windows logo testing, click Continue Anyway (Figure 6-7). Figure 6-7: Windows Logo Testing Window BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 75 6. At the Files Needed window (Figure 6-8), click Browse and select the CD-ROM drive. Figure 6-8: Files Needed Window 7. Double-click the Drivers folder (Figure 6-9). Figure 6-9: Locate File Window - Drivers REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 76 9. Click on the BiPAPFocus.inf file then click Open (Figure 6-11). Figure 6-11: Locate File Window - BiPAPFocus.inf 10. Use the file path suggested then click OK (Figure 6-12). Figure 6-12: Files Needed Window - File Path BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 77 Chapter 6 Ventilator Communications 11. If prompted to insert the Respironics CD (already in the drive), click OK (Figure 6-13). Figure 6-13: Insert Disk Window 12. At the Files Needed window, click Browse and select the CD-ROM drive (Figure 6-14).
  • Page 78 15. Click on the BiPAPFocus.sys file then click Open (Figure 6-17). Figure 6-17: Locate File Window - BiPAPFocus.sys 16. Use the file path suggested then click OK (Figure 6-18). Figure 6-18: Files Needed Window - File Path 6-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 79 18. Once the software installation is complete, click Finish to complete the installation (Figure 6-20). Figure 6-20: Found New Hardware Wizard Complete 19. Leave the ventilator in Download mode to begin ventilator software Installation. 6-11 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 80 1. With the software CD (P/N 1063919) into the CD-ROM drive, use on the Ventilator Windows Explorer to view the CD files (Figure 6-21). Figure 6-21: Files on the BiPAP Focus Software CD 6-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 81 2. Double-click Install_Focus_Software.exe to launch the BiPAP Focus Wizard (Figure 6-22), then click Next. Figure 6-22: BiPAP Focus Wizard 3. Select Install new BiPAP Focus Software (Figure 6-23), then click Next. Figure 6-23: Install New BiPAP Focus Controller Software 6-13 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual ©...
  • Page 82 6.1.) Click Next. Figure 6-24: Communications Cable Connection Screen 5. When Phase 1 BiPAP Focus Wizard appears, click Next (Figure 6-25). The software erases the controller PCB software (Figure 6-26), and then installs new software (Figure 6-27).
  • Page 83 Chapter 6 Ventilator Communications Figure 6-26: Erasing Controller PCB Software Figure 6-27: Reprogramming Controller PCB Software 6-15 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 84 (Figure 6-28). Figure 6-28: Phase 1 Complete, Start Phase 2 7. The Phase 2 BiPAP Focus Wizard opens to begin installing software on the user interface PCB. The ventilator should still be in Download mode and the cables connected. Click Next and follow the screen prompts (Figure 6-29).
  • Page 85 Figure 6-30: Option to View Cable Connection Instructions 9. The Wizard verifies that it can communicate with the ventilator. When the Next button is enabled click Next (Figure 6-31). Figure 6-31: Detecting Ventilator Connection 6-17 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 86 Ventilator Communications 10. The Wizard software then gathers the ventilator configuration information. Click Next (Figure 6-32). Figure 6-32: Searching for BiPAP Focus Ventilator window 11. When the ventilator configuration is displayed, click Next (Figure 6-33). Figure 6-33: Ventilator Configuration Window 6-18 BiPAP Focus Ventilator Service Manual ©...
  • Page 87 Figure 6-34: Selecting to Install new software 13. At available software window, select (in this example) Version 2.00 (File: D:\Loads\P2.00.004. TUI.scl) then click Next (Figure 6-35). Figure 6-35: Available Software Window 6-19 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 88 (Figure 6-37). Figure 6-37: Installing User Interface Software 6-20 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 89 Ventilator Communications 16. When software installation is complete click Next (Figure 6-38). Figure 6-38: Software Installation Complete 17. When BiPAP Focus Wizard Phase 2 Completion window appears, click Finish (Figure 6-39). Figure 6-39: BiPAP Focus Wizard Phase 2 Completion window 18.
  • Page 90: Reprogramming The Bipap Focus Serial Number

    Chapter 6 Ventilator Communications 6.4 Reprogramming The serial number of each BiPAP Focus Ventilator is stored in non-volatile memory on the user interface PCB, and must be reprogrammed if you replace the BiPAP Focus the PCB. Serial Number 1. Write down the serial number from the bottom of the unit, include hyphens if the serial number includes them.
  • Page 91 Chapter 6 Ventilator Communications 5. When the BiPAP Focus wizard asks what operation to perform (Figure 6-41), select I want to perform advanced configuration operations, then click Next. Figure 6-41: Selecting Advanced Configuration Operations 6. Select I want to modify the BiPAP Focus ventilator’s configuration, then click Next (Figure 6-42).
  • Page 92 Update Configuration button (Figure 6-43). Figure 6-43: Entering the Serial Number 8. When the configuration is successfully updated, click Next (Figure 6-44). Figure 6-44: Configuration Update Successful 6-24 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 93 Chapter 6 Ventilator Communications 9. At the BiPAP Focus Wizard Phase 2 Completion window appears, click Finish (Figure 6-45). Figure 6-45: Phase 2 Wizard Complete 10. The ventilator restarts in Normal mode using the default settings. Depending on the ventilator configuration, you may need to perform a calibration as described in Chapter 7.
  • Page 94 Chapter 6 Ventilator Communications (This page is intentionally blank.) 6-26 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 95: Testing, Calibration, And Performance Verification

    Follow manufacturers' labeling and institutional guidelines for disposal of contaminated accessories. Before servicing the ventilator, clean and inspect as described in periodic maintenance procedures (Chapter 4). REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 96: Test Equipment And Accessories

    Cable kit, BiPAP Focus communications Respironics P/N 1039800 (USB and serial cables) Tubing, silicone, 3/16-in. ID x 6.5 ft. Local supplier (for manometer connection) Flow control valve Respironics P/N 1006120 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 97: Recommended Testing

    Replacing the controller PCB and/or user interface PCB clears the calibration tables. Perform the system run-in once, between the two calibration procedures. (Do not perform system run-in before the first calibration.) REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 98: Ventilator Run-In

    RR = 4/min IPAP = 12 cmH I-Time = 1 sec EPAP = 4 cmH Ramp Time = 0 min Rise-Time = 2 Ramp Start = 4 cmH BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 99: Battery Test

    Chapter 7 Testing, Calibration, and Performance Verification 7.4 Battery Test WARNING: To avoid the risk of fire, use only Respironics-approved batteries. All battery connectors are keyed to ensure proper connection. Do not attempt to connect a battery incorrectly. NOTE: A Low Battery or Battery Depleted alarm during the pre-operational check indicates that the battery must be charged.
  • Page 100: Real-Time Clock Calibration

    Downloading Real-Time Clock Calibration Software Follow these steps to download the most current software to the service PC, then perform the real-time clock calibration: 1. From the service PC, go to http://my.respironics.com and click on Service Software (Figure 7-1). Figure 7-1: Select Service Software BiPAP Focus Ventilator Service Manual ©...
  • Page 101 3. At Service Center Tools Suite click Download (Figure 7-3). Figure 7-3: Service Center Suite 4. When the download is complete, extract the file Service-Center- Tools.exe and launch the installation wizard. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 102: Performing The Real-Time Clock Calibration

    The utility software requires you to use COM1 on the service PC. Do not connect or use the BiPAP Focus USB cable while connected to a docking station. Do not connect the BiPAP Focus USB cable through a USB hub.
  • Page 103 Figure 7-5: Selecting Set RT Clock 6. At the Real-Time Calibration Tool window, click the RT Clock to begin calibration (Figure 7-6). Figure 7-6: Real Time Clock Calibration Tool REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 104: If Real-Time Clock Calibration Fails

    If calibration fails, confirm that ventilator time and date are set to within 5 minutes of the service PC. Re-run the calibration. If the calibration fails again, replace the controller PCB and retest. 7-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 105: System Calibration

    Downloading System Calibration Software Follow these steps to download the most current calibration software to the service PC, then perform the calibration procedure: 1. From the service PC, go to http://my.respironics.com and click on Service Software (Figure 7-8). Figure 7-8: Select Service Software...
  • Page 106 Testing, Calibration, and Performance Verification 2. Click on Utility Tools (Figure 7-9). Figure 7-9: Select Utility Tools 3. At BiPAP Focus Calibration Software click Download (Figure 7-10). Figure 7-10: BiPAP Calibration Software 4. When the download is complete, extract the file BiPAP-Focus- Calibration_Software.exe and launch the installation wizard.
  • Page 107: Performing The System Calibration

    7.6.2 Performing the System Calibration The calibration procedure provides instructions for the remote calibration of the BiPAP Focus. The results of this calibration must be printed to a local printer. NOTE: The real-time clock calibration (section 7.5) must be complete before attempting the system calibration procedure.
  • Page 108 Pressure calibration: 7. Connect the oxygen enrichment adapter to the patient outlet of the ventilator, and use the silicone cork to occlude the adapter. 7-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 109 9. On the service PC, open the Service Calibration Software and start the calibration by clicking on the white Run arrow in the upper left corner of the window (Figure 7-13). arrow Figure 7-13: Starting Calibration 7-15 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 110 11. At the prompt, enter the ventilator model number and click OK. 12. Enter operator name and click OK. 13. Verify that the label is BiPAP Focus and click PASS. 14. The program runs an alarm test: follow the onscreen instructions.
  • Page 111 Figure 7-16: Entering Pressure Reading 18. The program prompts you to calibrate at various pressure levels using the onscreen buttons to increase or decrease blower speed to achieve the correct pressures. 7-17 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 112 Connect another length of patient tubing from the flowmeter outlet of the to the control valve (completely open the flow valve) (Figure 7-18). Figure 7-18: Positive Flow Calibration Measurement Setup 7-18 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 113 146 LPM, then click OK. 24. The screen then prompts you to adjust the flow valve to achieve various flow settings. Ensure that each flow setting is achieved before clicking OK. 7-19 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 114 For the negative flow source, a REMstar device may be used with the pressure set at 20 cmH Air flow into ventilator from external negative flow source Figure 7-21: System Calibration With Negative Flow Source 7-20 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 115 29. At the prompts, adjust the flow control valve to achieve various negative flows, enter the flowmeter readings, then click OK. 30. At the Remove Negative Flow Resource prompt, remove the negative flow source and click OK. 7-21 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 116 Adobe PDF) or a printer connected to the PC. If a print error occurs at the end of the calibration, the calibration table is cleared and the calibration must be repeated. Figure 7-24: BiPAP Focus Test Results 7-22 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 117 Chapter 7 Testing, Calibration, and Performance Verification Figure 7-25: Print Error - No Printer Connected 7-23 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 118: Electrical Safety Test

    EMC purposes only. The Protective Earth system is within the power supply unit as it has no conductive exposed parts. The accessible functional earth is allowed by IEC 60601-1 because the BiPAP Focus meets enclosure leakage and separation requirements. It should be noted that the ground impedance test per IEC 601 is a type test and not a production line or routine safety test.
  • Page 119: Alarm/Led Test

    8. Remove the DMM leads and return to the System Test screen. 9. Select Alarm LED On, then Silence LED On, and then On Battery LED On to test LEDs. 7-25 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 120: Pressure Accuracy Test

    O, EPAP = 5 cmH O, Rate = 10 BPM, and I-Time = 2.0. 7. Verify that the BiPAP Focus switches between IPAP and EPAP. Record the results. 8. Activate the ramp time setting and set ramp time to 5 minutes.
  • Page 121: Performance Verification Data Form

    Forward leakage current: < 100 microamperes (100-120 VAC) < 300 microamperes (230-240 VAC) Reverse leakage current: < 300 microamperes (100-120 VAC) < 500 microamperes (230-240 VAC) 7-27 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 122 O, measured pressure = 4.5-5.5 cmH When IPAP = 10 cmH O, measured pressure = 9-11 cmH When EPAP = 5 cmH O, measured pressure = 4.5-5.5 cmH after 5 minutes 7-28 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 123: Component Removal/Installation

    When reconnecting the tubing to the controller PCB sensors, ensure that the tubing fits tightly. If the tubing has become loose, clip a short length (2 mm) from the end of the tubing to ensure a snug fit. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 124: Disconnecting Power

    3. Remove the battery cover and battery pack. 4. Disconnect the battery Battery compartment Screws (x2) (5 lbf-in., 57 N cm) Battery cable connector Figure 8-1: Disconnecting Battery Power BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 125: Top Cover

    (5 lbf-in., 57 N cm) Side panel Display assembly Screws (x2) (8 lbf-in., 90 N cm) Side panel Screws (x2) (5 lbf-in., 57 N cm) Figure 8-2: Bottom Cover REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 126 Jack screws (2) ON/OFF switch bezel Figure 8-4: Bipap Focus Back Panel 5. Tilt the bottom of the display assembly outward to separate it from the bottom and top covers. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 127 Top cover Screws (x4) (8 lbf-in., 90 N cm) Bottom cover Figure 8-5: Top and Bottom Covers REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 128: Display Assembly

    3. Disconnect the ground cable from the user interface PCB. User interface PCB Display assembly UI/VGA ribbon cable UI/VGA ground cable Figure 8-6: Display Assembly and User Interface PCB BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 129 8. Pull the ground cable through the rear display cover and completely open the display assembly. 9. Remove the 4 screws that secure the VGA LCD and VGA interface PCB. 10. Disconnect the cables from the VGA interface PCB. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 130 11. Replace the display assembly or its replaceable components as necessary. NOTE: Tuck ribbon cables under VGA interface PCB before securing screws. Do not twist or pinch cables between covers. BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 131: User Interface Pcb

    • Battery cable (from battery housing) • Interconnect power cable (controller PCB) and retaining screw • Interprocessor communications cable 3. Remove the 4 screws from the spacer nuts. REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 132: Controller Pcb

    Spacer nuts (x3) Nylon washers (x3) Controller Low-profile service clip Interconnect power cable Figure 8-11: Controller PCB 5. Lift the PCB to expose the remaining cables and tubing. 8-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 133 If the tubing has become loose, clip a short length (2 mm) from the end of the tubing to ensure a snug fit. 8-11 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 134: Flow Valve

    6. Disconnect the diffuser from the flow valve. 7. Remove or replace the flow valve. 8-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 135: Blower

    3. Remove the controller PCB (section 8.5). Screws (x2) (5 lbf-in., 57 N cm) Diffuser Blower Figure 8-14: Blower 4. Remove the 2 screws that anchor the blower to the muffler. 8-13 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 136 Align the blower screw holes with the screw holes in the muffler. 8-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 137: Backup Alarm

    5. Remove the 2 screws that attach the backup alarm. 6. Remove the backup alarm and foam gasket. Screws (x2) (3 lbf-in., 34 N cm) Backup Diffuser alarm Figure 8-16: Backup Alarm 8-15 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 138: Primary Alarm

    Primary User connector alarm interface Figure 8-17: Primary Alarm 2. Disconnect the primary alarm connector from the user interface PCB. 3. Remove the 2 primary alarm screws. 8-16 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 139: Bottom Cover And Muffler

    Screws (x4) 8 lbf-in., 90 N cm Low-profile service clip Foam Muffler Bottom cover Bottom cover with Bottom cover with muffler installed muffler removed Figure 8-18: Bottom Cover 8-17 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 140: Interconnect Diagrams

    Chapter 8 Component Removal/Installation 8.11 Interconnect This section shows power (Figure 8-19) and communication (Figure 8-20) interconnections. Diagrams External AC/DC power supply Figure 8-19: Power Interconnection 8-18 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 141 Chapter 8 Component Removal/Installation The BiPAP Focus ventilator has the following communications requirements: • Interprocessor communication (IPC) between the controller PCB and the user interface PCB. • Serial port for calibration and controller PCB software upgrade. • USB port for user interface PCB software upgrade.
  • Page 142 Chapter 8 Component Removal/Installation (This page is intentionally blank.) 8-20 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 143: Bipap Focus Replacement Parts List

    Chapter 9. BiPAP Focus Replacement Parts List This chapter lists and illustrates the major components of the BiPAP Focus ventilator, including: • Exterior • Front panel/user interface • Ventilator interior • Electronic parts NOTE: When replacing the bottom cover, order the main cover assembly (P/N 1052538) and a main label (P/N 1031229).
  • Page 144 Chapter 9 BiPAP Focus Replacement Parts List Table 9-1: Complete BiPAP Focus Service Parts List (Continued) Description Quantity Part number Cover assembly, display, BiPAP Focus 1052539 Cover, display front Cover, display rear Cover, display side Cover assembly, main, BiPAP Focus...
  • Page 145 Chapter 9 BiPAP Focus Replacement Parts List Table 9-1: Complete BiPAP Focus Service Parts List (Continued) Description Quantity Part number Power cord, Italy 1034953 Power cord, Switzerland 1034955 Power cord, UK/Ireland 1029982 Power cord, USA 1029989 Power supply, AC/DC, 120 W, 18 V...
  • Page 146 • Rear access door • Battery cover Side cover Black rubber foot (x4) P/N 1031227 M3 x 8 screws (x4) P/N 1030148 Battery cover Rear access door P/N 1025761 P/N 1025766 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 147 ON/OFF switch bezel Power cord strain relief Main label P/N 1031160 clip P/N 1031229 P/N 1031223 Reusable particulate Disposable particulate filter (gray) filter (white) P/N 1005964 P/N 1005945 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 148 Chapter 9 BiPAP Focus Replacement Parts List 9.3 Front Panel/User Interface VGA interface PCB VGA LCD P/N 1012280 P/N 1012282 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 149: Ventilator Interior

    P/N 1062043 Blower assembly P/N 1023818 Flow valve seal, foam P/N 1011998 Grommet (x3) Flow valve P/N 1017912 Flow valve strap P/N 1031236 Blower gasket, silicone P/N 1056851 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 150 Chapter 9 BiPAP Focus Replacement Parts List Backup alarm Primary alarm P/N 1039803 P/N 1041089 Muffler assembly P/N 1017913 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 151: Communications Cables

    Chapter 9 BiPAP Focus Replacement Parts List 9.5 Communications Cables BiPAP Focus communications cable kit P/N 1039800 Serial cable USB cable REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 152: Electronic Parts

    P/N 1025775 P/N 1028006 Power cord Denmark P/N 1029991 Europe P/N 1029985 India/RSA P/N 1034954 Italy P/N 1034953 Switzerland P/N 1034955 UK/Ireland P/N 1029982 USA P/N 1029989 9-10 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 153 P/N 1025776 Switch control cable P/N 1031159 UI/VGA ribbon cable P/N 1031239 UI/VGA GND cable P/N 1040723 Low-profile service clip (plastic, adhesive- backed, ribbon cable) P/N 1030057 9-11 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 154 Chapter 9 BiPAP Focus Replacement Parts List (This page is intentionally blank.) 9-12 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 155: Specifications

    IPAP cannot be set below EPAP. EPAP Range: 4-25 cmH O (4-25 hPa) Resolution: 1 cmH O (1 hPa) Dynamic accuracy: ± 5 cmH O (5 hPa) 10-1 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 156 60 L/min is less than 1 cmH O duet to mask and ventilator design. Rate Range: 0-60/min Resolution: 1/min Accuracy: ± (1 + 10% reading) 10-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 157 Range: 20, 40, 60 sec or OFF Default setting: 20 sec #Apnea Range: 5, 10, 20, or OFF NOTE: For the first hour, this value is an estimate Default setting: OFF only. 10-3 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 158 • Female DB9 connector pin configuration: WARNING: Use only Respironics-supplied cables with the nurse call station/serial communications connector. 10-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 159 24 VAC or 36 VDC at 500 mA, with a minimum current of 1 mA. Fuse type No replaceable fuses. Power cord Use the BiPAP Focus power cords listed in Chapter 9. Physical Ventilator 355 mm (H) x 290 mm (W) x 140 mm (D) dimensions 14.0 in.
  • Page 160: Patient Circuit Details

    Chapter 10 Specifications 10.2 Patient Circuit The BiPAP Focus System automatically compensates for pressure drops associated with a Vision single patient use circuit and the mandatory bacteria Details filter. The pressure drop increases as restrictive elements (for example, humidifiers) are added to the circuit. Always use a manometer to verify patient mask pressure.
  • Page 161: Maximum Pressure Drop For Disconnect

    When the ventilator is not in use for periods of 2 weeks or longer, it is recommended that the unit be stored in a clean, sanitized plastic bag. 10-7 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 162: Date Of Manufacture

    This serial number indicates a manufacture date of 29 July 2006: S/N 8772 060729-01 P/N 1028772 Unit number Date: YY/MM/DD Include the hyphen if present (varies by build date) when recording the serial number 10-8 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 163: Compliance

    Guidance and manufacturer’s declaration: electromagnetic emissions, IEC 60601-1-2 + A1:2000. The BiPAP Focus System is suitable for use in the specified electromagnetic environment. The customer and/or the user of the BiPAP Focus System should ensure that it is used in an electromagnetic environment as described below:...
  • Page 164 Guidance and manufacturer’s declaration: electromagnetic immunity, IEC 60601-1-2 + A1:2000. The BiPAP Focus System is suitable for use in the specified electromagnetic environment. The customer and/or the user of the BiPAP Focus System should ensure that it is used in an electromagnetic environment as described below:...
  • Page 165 IEC 61000-4-8 commercial and/or hospital environment. NOTE: UT is the AC mains voltage prior to application of the test level. Guidance and manufacturer’s declaration: electromagnetic immunity, IEC 60601-1-2. 10-11 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 166 To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the BiPAP Focus System is used exceeds the applicable RF compliance level above, the BiPAP Focus System should be observed to verify normal operation.
  • Page 167: Warranty

    10.7 Warranty Respironics warrants the BiPAP Focus Ventilator to be free from defects in material and workmanship for a period of 12 months from the date of purchase, provided that the unit is operated under conditions of normal use as described in this manual.
  • Page 168 Unauthorized service and/or failure to perform periodic maintenance may void this warranty. This warranty does not cover damage that may occur in shipment. This warranty does not cover internal battery replacement. 10-14 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 169 4-5 Date and time, setting, 5-4 Battery cable Date of manufacture, 5-3, 10-8 part number, 9-11 Default settings, 7-4, 10-1 Diagnostic mode date and time, 5-4 Index-1 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 170 Flow and pressure accuracy, 7-22 Notes, cautions, and warnings, 2-1 Flow valve Nurse call/serial port, 3-8 description, 3-3 pinout, 10-4 part number, 9-7 removal/installation, 8-12 ON/OFF switch bezel part number, 9-5 Index-2 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...
  • Page 171 S/T mode settings, 10-1 downloading software, 7-6 warranty, 10-13 performing calibration, 7-8 Storage, 4-9 Switch control cable part number, 9-11 System, 7-11 System calibration, 7-11 downloading software, 7-11 performing, 7-13 Index-3 REF 1029568 Rev B BiPAP Focus Ventilator Service Manual © Respironics, Inc.
  • Page 172 9-10 removal/installation, 8-9 Ventilation modes, 3-1 Ventilator storage, 4-9 VGA display description, 3-7 VGA interface PCB description, 3-7 part number, 9-6 removal/installation, 8-6 VGA LCD removal/installation, 8-6 Index-4 BiPAP Focus Ventilator Service Manual © Respironics, Inc. REF 1029568 Rev B...

Table of Contents