8.10 Bottom Cover
and Muffler
Muffler
Bottom
cover
Bottom cover with
muffler installed
REF 1029568 Rev B
Follow these steps to remove the bottom cover and muffler (Figure 8-18).
Reverse to install.
1. Remove top cover (section 8.2).
2. Remove the user interface PCB (section 8.4).
3. Remove the controller PCB (section 8.5).
4. Remove the flow valve (section 8.6) and the outlet port, air flow
sensor, silicone connector, and diffuser.
5. Remove blower (section 8.7).
6. Remove the four screws from the muffler that anchor it to the bottom
cover.
7. Remove the foam if necessary, and clean any adhesive residue with
isopropyl alcohol. Install new foam if needed, and apply product
labeling if replacing the bottom cover.
Screws (x4)
8 lbf-in., 90 N cm
Low-profile service
clip
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Component Removal/Installation
Foam
Bottom cover with
muffler removed
Figure 8-18: Bottom Cover
Chapter 8
8-17