REF 1029568 Rev B
6. Remove the four screws holding the top and bottom covers together (8
lbf-in., 90 N cm). Lift front edge of top cover and remove, ensuring
that the back panel cut-outs clear the serial port and switch.
Bottom cover
Figure 8-5: Top and Bottom Covers
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
Component Removal/Installation
Top cover
Chapter 8
Screws (x4)
(8 lbf-in., 90 N cm)
8-5