ZW-8000/7000/5000 Series can hold up to eight sets of sensing settings, which are called "bank (BANK)".
When the setup is changed, the bank can be switched externally.
What is Bank?
The sensing settings for measurement
Sensing
Output
Example of switching bank for settings
If you register settings of various classes,
Sensing setting for product class A
Sensing
Output
Measuring operation
with the contents set
in BANK1
Important
• If you want to register multiple bank data having the same values set except for "threshold values (judgment
values)", by changing the mode of the bank, you can increase the number of banks from 8 to 32.
Setting the Bank Mode p.188
• The "output destination" (current output value/voltage output value) in the I/O setting parameters is set the same for
all banks. The output destination cannot be set separately for individual banks.
"2-1 Parallel I/O connection" described in Displacement Sensor ZW-8000/7000/5000 series Confocal
Fiber Type Displacement Sensor User's Manual for Communications Settings (Z363)
28
Basic Operation Flow
Filter
I/O
Threshold value
Filter
I/O
Threshold value
BANK1
BANK2
BANK3
BANK8
Sensing
TASK1
Measurement point
Filter
Threshold value
Output
I/O
are held as one bank.
Sensing
Output
Register the different settings
for changing the device setup.
Sensing setting for product class B
Sensing
Output
Switch
to BANK2.
It is easy to change the device setup.
TASK4
Measurement point
Filter
Threshold value
Output
BANK1
Filter
I/O
Threshold value
BANK
Measuring operation
with the contents set in
BANK2
Filter
I/O
Threshold value
BANK1
BANK2
BANK3
BANK8
TASK1 to TASK4 are
specified in the settings
for each bank.