Confocal fiber type displacement sensor
Table of Contents

Advertisement

Multi-point scaling
With ZW-8000@, measurements are taken at multiple locations and a precision stage is used to minimize the
measurement error between the measurement value and the actual distance.
This function is enabled when nonlinearity occurs in the measurement error of the measurement value and the
actual distance.
This function cannot be used with ZW-7000@/ZW-5000@.
Example: When correcting measurement values to match actual distances at 4 points
Actual distances
(mm)
Item
Setting item
Scaling
Scaling points
(Multi-point
(n)
scaling)
Set value (1st
point)
Set value (Point2
to Point(n-1))
Measurement
value (Point(n))
Important
Separate each distance between the N specified points by at least 1% of the rated measuring range for the connected
Sensor Head.
For example, for the ZW-S8010, the two measured points must be separated by at least 1 mm × 0.01 = 0.01 mm as
the measuring range is 1 mm (± 0.5 mm).
148
Setting the Output Conditions
Measured value
(mm)
Setting value
2 to 10 [point]
-999.999999 to 999.999999 [mm]
-999.999999 to 999.999999 [mm]
-999.999999 to 999.999999 [mm]
Description
Set the points on which to execute scaling.
Correct the measured value of Point1, and set the ideal
value to display.
Correct the measured value of Point2 to Point(n-1), and set
the ideal value to display
Correct the measured value of Point(n), and set the ideal
value to display.