Setting Triggers
Set how measurement start to end timing is to be input.
Item
Setting item
Setting value
Trigger
Trigger method
External input
Set-
tings
Selfup trigger
Selfdown trigger
ZW-8000/7000/5000
User's Manual
Description
Enter the trigger for the start of sampling in "TIMING input" for the 32-pole
terminal block.
The period that the signal input to "TIMING input" is ON is the sampling period.
The measurement values actually sampled vary according to the TIMING
input mode setting.
ON
Timing input
OFF
Important
When a delay time is set, input OFF and sampling period end are not
synchronized. End timing is after the specified sampling period has elapsed.
The period in which the measured value exceeds the preset self-trigger level
is taken to be the sampling period.
Hold measurement can be performed without a synchronous input.
Self-trigger
level
Measured
value
Sampling period
Important
When a delay time is set, the timing that the measured value falls
below the self-trigger level and the sampling period end are not
synchronized. End timing is after the sampling period has elapsed.
The period in which the measured value falls below the preset self-trigger
level is taken to be the sampling period.
Hold measurement can be performed without a synchronous input.
Measured
value
Self-trigger
level
Sampling period
Important
When a delay time is set, the timing that the measured value
exceeds the self-trigger level and the sampling period end are not
synchronized. End timing is after the sampling period has elapsed.
Sampling period
Hysteresis width
(for self-trigger)
Action point
Return point
Sampling period
Hysteresis width
(for self-trigger)
Action point
Return point
Setting the Output Conditions
163