Arrange the specified current values (measurement values) of Point2 to Point(n) so that they are in ascending or
descending order.
An error will occur if this relationship is not adhered to.
Point1
Point2
·
·
Point(n)
Note
The current measured value is displayed at [Current value], and is refreshed at 100 ms cycles. When [STOP] is
clicked, refreshing of the [Current value] display stops. Clicking the [STOP] button again resumes display refreshing.
5
Click [OK].
Note
• Multi-point scaling can also be executed by the operating keys on the Sensor Controller.
Multi-point scaling (ZW-8000@ only) p.249
• The set correction values can be exported by clicking [Numerous Sections Adjustment Data Export]. Also, any
previously exported numerous sections adjustment data can be imported by clicking [Numerous Sections
Adjustment Data Import].