Omron ZW-8000 Series User Manual page 163

Confocal fiber type displacement sensor
Table of Contents

Advertisement

Item
Setting item
Hold
Hold mode
 Multi View Explore
→ Edit pane
→ Task Settings window : [Hold]
1
Select the hold mode from [Hold mode].
Setting items for required parameters are
displayed depending on the selected hold
mode.
Setting Triggers p.163
Setting a Trigger Delay p.167
Note
The hold mode can also be set by the operating keys on the Sensor Controller.
Setting the Hold Mode p.251
ZW-8000/7000/5000
User's Manual
Setting value
Description
Auto Peak to
Holds the difference between the maximum and minimum values of the
Peak
measurement result. Output the measurement value at each updating of the
difference value Maximum and Minimum.
0 0
Average
Holds the "average of the measured value" of the sampling period. The current
measurement value is output after Sampling, and then this state is held until
the next sampling ends.
Current
measured value
Sampling
Holds the measured value the moment that control enters the sampling
period. The current measurement value is output after Sampling, and then this
state is held until the next sampling is started.
Current
measured value
Through
The hold function is not used. The measured value is always output.
(default value)
: [Bank Group] | [(Bank Data Name)] (double click)
: [Task Settings] icon (
Output
(Max.value - Min.value)
Current
measured value
Sampling period
Sampling period
)
Setting the Output Conditions
Output
(Average)
Output
161

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

Zw-7000 seriesZw-5000 series

Table of Contents