9- Technical Specifications - Saadat Respina P1 User Manual

Table of Contents

Advertisement

9- Technical Specifications

Technical Specifications
Assisted Control Mandatory Ventilation
Ventilation
Synchronized Intermittent Mandatory Ventilation
Modes
Spontaneous Ventilation
Volume–controlled breaths
Pressure–controlled breaths
Breath Types
Volume Targeted Pressure–controlled breaths
(Pressure Regulated Volume Control) & (Volume
Support)
Dual level PEEP breaths
Patient Height
Patient Type
IBW (Ideal Body Weight )
Gender
Compliance
VBS (Ventilator
Breathing
Inspiratory Resistance
System)
Expiratory Resistance
Apnea Backup
Settings
Pressure triggering
Breath triggering
Flow triggering
Respiratory Rate
Accuracy: ±1 b/min or 4% wig
Respiratory Rate apnea
Accuracy:±1 b/min or 4% wig
Tidal Volume
Accuracy:
For 50–2000 ml: ± (4 ml + 15% Actual reading)
For 20–49 ml:
PEEP / CPAP
Accuracy: ±12% or ±(2cmH2O +4% Actual reading)
wig
Pcontrol
Accuracy: ±12% or ±(2cmH2O +4% Actual reading)
wig
Additional
Setting
Psupport
Accuracy: ±12% or ±(2cmH2O +4% Actual reading)
wig
Peak Flow
Accuracy: ±12% or ±(3 l/min +4% Actual reading)
wig
Flow Pattern
NIV
Non- Invasive Ventilation
Leak Comp
Automatic Leak Compensation
I–Time (Ti)
Accuracy: ± 0.02 sec
± (7 ml + 15% Actual reading)
A/C
SIMV
SPONT
VCV, V–SIMV (With Dual control
in the breath)
PCV, P–SIMV, PSV
PRVC–CMV, PRVC–SIMV, VSV
APRV (With Auto Tlow)
42 – 250 cm
5 – 200 Kg
Male, Female
0.5 – 5 ml/cmH2O
0 – 2.5 cmH2O.s/l
0 – 2.5 cmH2O.s/l
On (Rate, Pcontrol, Ti); OFF
(-2) – (-20) cmH2O
2 – 20 l/min
2– 150 b/min
5 – 150 b/min
20 – 2600 ml
(Compliance & BTPS compensated)
0 – 50 cmH2O
2 – 80 cmH2O
0 – 80 cmH2O
0.2 – 250 l/min
Square, Decel.
On or Off;
Available in P-SIMV, PSV and apnea
modes.
On or Off
0.1 – 10.0 sec
92