4-3

100% O2 Key

Pressing 100% O2 key on the front panel (if the oxygen source is available with proper
pressure) will deliver a mandatory or assisted breath (like applied pattern in Manual
inspiration) depending on ventilation mode and time of key pressing and 100% oxygen will
be delivered for 2 minutes in the current ventilation mode. After 2 minutes the delivered
oxygen will return to its original setting.
"100% O2" is shown on the screen throughout the oxygen delivery.
If you press O2 100% key while the oxygen source is not available or disconnected, "100%
O2 Not Available" alarm will be activated and after supplying the oxygen, the maneuver will
resume.
If you press the Standby key, the maneuver will stop. The maneuver will resume if you exit
the system from the standby mode. This maneuver cannot be enabled in the standby mode.
If mode, setting parameters and other maneuvers are changed, O2 100% maneuver will be
disabled.
4-4

Manual Key

Press this key on the ventilator front panel to deliver one breath to patient in the current
setting. In this condition "Manual" is displayed on the screen. In A/C and APRV modes the
manual inspiration is mandatory and in PSV and VSV modes the manual inspiration is an
assisted breath. In SIMV modes the manual inspiration could be mandatory or assisted breath
depending on maneuver. The manual maneuver is active in all ventilation modes.
It should be noted that if the device is set so that the exhalation time is less than 750
milliseconds, Manual breath will not apply.
When the Manual key is pressed one breath is delivered during the specified interval, for the
next breaths the key should be pressed again (i.e. every key pressing will deliver one manual
breath).
You cannot enable or disable this maneuver in the Standby mode.