Saadat Respina P1 User Manual page 96

Table of Contents

Advertisement

Active Exhalation valve (Pressure relief during
inhalation)
Nebulizer
Configuration
Setting
Humidification Type
Compliance Comp
AC input
Battery Backup (with fully charged batteries)
High pressure air inlet supply
Power and Gas
Supply
High pressure oxygen inlet supply
Battery
Type
specification
Voltage, Energy
Auditory alarm
Visual alarm
Alarm System
Indicator
Types
Priorities
Operating temperature
Environmental
Data
Storage and transport temperature
Operating Humidity
Storage and transport Humidity
Operating altitude
Width x depth x height (main case)
Physical Data
Weight
Noise Level
Display
TFT Size
Maximum limited pressure
Pressure Safety
Maximum operating pressure
Classifications
Compliance and
Approvals
Meets International Standards
Available in all modes
Duration: 1-480 min, Resolution:1
Accuracy: ±10 s
None, HME, Humidifier
On or Off;
100 to 240 VAC (47 – 63 Hz)
Up to 5 hours
Clean oil-free and medical grade
2.4 – 6.0 bar, full performance
1.0 – 2.4 bar, degraded performance
0.0 – 0.8 bar, alarm
Clean dry and oil–free medical grade
oxygen
2.4 – 6.0 bar, full performance
1.0 – 2.4 bar, degraded performance
0.0 – 0.8 bar, alarm
Two separate Li- Ion batteries
11.1 V, Totally> 48 Wh
Internal speakers
In the TFT display
LED
Technical, Physiological, Messages
High, Medium, Low
Battery operation: 5 – 40 °C
Mains operation, battery full charged:
5 – 35 °C
Mains operation, battery charging: 5
– 30 °C
(-10) – 60 °C
15% – 95% non condensing
5% – 95% non condensing
11,000 ft (3,500 m) above sea level
45 x 35 x 29 cm
10.2 Kg
TBD
12.1" wide, with Touch screen
110 cmH2O
80 cmH2O
Class I, Type B, internally powered,
drip–proof equipment for continuous
operation
EMC related group and
classification: I-B
EN 60601-1: 2006
EN 60601-1-2: 2015
EN 60601-1-8: 2007/ AC: 2010
ISO 80601-2-12: 2020
96