High O2 Pressure
Low O2 Pressure
High Air Pressure
Low Air Pressure
High Gas
Temperature
Error code 17
Error code 2
Fatal Error 5
Fatal Error 6
Medium Priority
Alarm
High Rate
Low Rate
Vti Limit Reached
High Vte
The oxygen inlet pressure is more
than the specified level.
The oxygen inlet pressure is lower
than the specified level.
This alarm will not be enabled if the
oxygen concentration is set to 21%.
The air inlet pressure is more than
the specified level.
The air inlet pressure is lower than
the specified level.
This alarm will not be enabled if the
oxygen concentration is set to 100%.
If the measured gas temperature
parameter (gas temp) exceeds the
allowed limit (45 ° C).
--
--
--
--
Alarm Definition
Respiratory rate (including
spontaneous and mandatory breaths)
has exceeded high alarm setting in
three respiratory cycles.
Respiratory rate (including
spontaneous and mandatory breaths)
has violated low alarm setting in
three respiratory cycles.
Inspired tidal volume has exceeded
high alarm setting in three
consecutive respiratory cycles.
Exhaled tidal volume has exceeded
high Vt
alarm setting in three
e
consecutive respiratory cycles.
Ventilation continues.
If the Air source is available,
ventilation continues with Air only.
The safety valve opens if the pressure
of both Air and Oxygen sources drops
or disconnects.
Ventilation continues.
If the Oxygen source is available,
ventilation continues with Oxygen
only.
The safety valve opens if the pressure
of both Air and Oxygen sources drops
or disconnects.
Ventilation continues.
Ventilation continues.
Ventilation continues.
Ventilation continues with the
following settings: Mode: PCV, Rate:
15, Ti: 1, PEEP: 2, Pcontrol: 20,
Trigger: Pressure, Ptrig: -3, O2: 100
Ventilation continues with the
following setting: Mode: PCV, Rate:
15, Ti: 1, PEEP: 2, Pcontrol: 15,
Trigger: Pressure, Ptrig: -3, O2: 100
Ventilator reaction to alarm
Ventilation continues.
Ventilation continues.
The inhalation phase will switch to
the exhalation phase (whether alarm
is activated or not) once Vti has
exceeded alarm limit.
Ventilation continues.
51