Warning
Incorrect setting of alarm limits (manually) could affect functionality of the
alarm system.
Alarm setting of different parameters is provided in below table:
Alarm limit
PEEP high
PEEP low
Rate high
Rate low
MV high
MV low
Vte high
Vte low
Figure 5-10 Alarms Menu
Parameter
MV High
MV Low
Vte High
Vte Low
Rate High
Rate Low
Equation
PEEP
monitoring
PEEP
monitoring
1.5
*
Rate
0.5
*
Rate
2
*
MV
0.5
*
MV
2
*
Vte
0.5
*
Vte
Alarm limits
[0.2- 50]
[0.1- 49.5] or Off
]
-
[
10
3000
or Off
[5 - 2995] or Off
]
[
3-180
]
[
2-179
+ 5
-
5
monitoring
monitoring
monitoring
monitoring
monitoring
monitoring
Unit
l/min
l/min
ml
ml
b/min
b/min
35