Cleaning And Disinfecting - Saadat Respina P1 User Manual

Table of Contents

Advertisement

Exhalation Valve
After each patient or as required, clean the exhalation valve by rinsing and drying with soap
and water. And then disinfect or sterilize.
Exhalation valve- flow sensor interface
After each patient or as required, clean the exhalation valve by rinsing and drying with soap
and water. And then disinfect or sterilize.
Warning
Replace faulty exhalation flow sensor.
Replace expiratory filter and HME every 24 hours or as needed. It should be
noted that expiratory filter and HME are disposable and must be replaced for
each patient.
Replace faulty exhalation membrane. Note that the exhalation membrane is
disposable and must be replaced for each patient.
7-3

Cleaning and disinfecting

 Before disinfecting the ventilator, make sure that the equipment is switched off and
disconnected from the power line.
 Exterior surface of the device should be disinfected after each use. Use isopropyl
alcohol to disinfect.
 Do not use strong solvents such as acetone or ammonia.
 Do not use rough material, such as steel wool, etc.
 Do not let the cleaning agent enter into the housing of the system.
 Dry out the cleaning agents on any part of the device.
62