Alarm Message
AC Unplug
Battery Defect
No Battery
Error code XX
Fatal Error X
100% O2 Not
Usable
Neb. Not Usable
Suction Not Usable
Possible Cause
The device is unplugged.
The battery failure
The internal battery is not available.
--
--
High pressure oxygen source (higher
than 1,2 bar) is not available.
Flow setting is lower than 11L/min.
High pressure oxygen source is not
available.
The oxygen source pressure is out of
range and the nebulizer will not
activate.
-The device is in the NIV mode.
- 100% O2 key is pressed before
selecting the suction maneuver.
- "Low O2 Pressure" alarm is active.
Corrective Action
- Plug in the device.
Contact the Customer Service
department.
- Ensure that a correct battery has
been inserted in the device.
- If the alarm persists, contact the
Customer Service department.
- Contact the Customer Service
department.
- Contact the Customer Service
department.
Check the inlet oxygen source
pressure.
Check that flow is not higher than
11L/min.
Check the inlet oxygen source
pressure.
- Disable the NIV mode.
- Stop 100% O2 phse.
- Check pressure of the oxygen
source.