If New from Patient Setup menu is selected, the event log will be cleared and after IBW
setting subsequent events will be stored. If Previous is selected from Setup menu, new events
will be added to the current log.
By clicking on each event, you can view current settings of time of alarm occurrence below
the Event Log.
The adjusted value is shown in Set column and the measurement value triggered alarm is
shown in Monitor column.
All events in the Event Log include the following data:
 Settings
 Alarm event as well as setting and monitoring values of relevant parameter
 Adjusted high and low alarm limits (using AutoSet key or manually)
 Config settings of the device including:
Nebulizer, Humidification Type, Compliance Compensation
 Placing the ventilator in standby mode and reactivating ventilation mode
 Turning off (software or hardware) and turning on the device
 Possible faults with the memory of the device reported in self-test
Note:
Power supply interruption would have no effect on Event Log content.
5-8

Config.

The following configurations could be performed in this menu:
Figure 5-11 Event log
37