Exh Not Fitted
O2 Sensor Error
Inverse Ratio
Neb. Not Usable
100% O2 Not
Usable
High Gas
Temperature
Suction Not Available
Error code 1
Error code 3
Error code 7
Error code 8
Exhalation flow sensor is not placed
correctly.
O2 sensor failure
Exhalation time is more than
inhalation time.
The nebulizer does not function.
100% O2 key does not function.
If the measured gas temperature
parameter (gas temp) exceeds the
allowed limit (42 ° C).
The suction maneuver could not be
performed.
--
--
--
--
Some parameters such as Vte that are
measured by this sensor will be
displayed blank.
Ventilation continues.
If flow trigger mode is active, the
device will switch to a pressure
trigger mode. In this condition you
cannot select flow trigger mode.
Some parameters such as Vte that are
measured by this sensor will be
displayed blank.
Ventilation continues.
O2 parameter is displayed blank.
Ventilation continues.
The nebulizer is disabled. When the
nebulization condition is provided
(flow is higher than 11L/min and
pressure of the oxygen source is 1,2
bar), the nebulizer will be activated.
100% O2 maneuver is disabled. When
the oxygen source is available, 100%
O2 is delivered.
Ventilation continues.
The maneuver is disabled.
Ventilation continues.
Ventilation continues.
Ventilation continues.
Ventilation continues.