Apnea Time
The default values of alarm limits are listed below:
Alarm limit
MV Low
Vte Low
Rate Low
PEEP Low
Leak
Vti Limit
Note:
1 . Setting the minimum of Ppeak alarm parameter depends on setting.
2 . High Ppeak alarm limit is set manualy only.
3 . In case of power failure and unavailability of the battery, alarm settings will be saved.
4 . The lower limit of MV alarm can be set to Off only in NIV ventilation. If the user sets
the alarm limit to Off, this value changes to 0.1 when exiting non-invasive breathing.
5 . The default of upper and lower lomits of Vte alarm and the lower limit of PEEP alarm
are Off.
6 . Alarm limits which are set to Off by the user, will not change by pressing Autoset
key.
Event Log
By selecting Event Log from Monitoring menu, you can view the latest alarm events and
settings in order of occurrence (up to 2000 events). when the event log list is filling, to record
new event, the oldest one is omit from the list, and new event is record as the 2000th event.