Guidance and manufacturer's declaration – electromagnetic immunity
The Ventilator Respina P1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the Ventilator Respina P1should assures that it is used in such an
environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Proximity fields from RF
wireless
communications
equipment
IEC 61000-4-3
Port
3 V
Input a.c. power
0,15 MHz – 80 MHz
6 V
in ISM bands
PATIENT coupling
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
Signal input/output
parts
3 V/m
80 MHz – 2,7 GHz
ENCLOSURE
80 % AM at 1 kHz
Refer to the following table
ENCLOSURE
(table 9 of EN 60601-1-2:
2015)
Compliance level
Electromagnetic
environment –
guidance