Saadat Respina P1 User Manual page 59

Table of Contents

Advertisement

Figure 7-2 O2 Sensor Calibration page
 System Test
The system test should always be performed before using the ventilator or after replacing the
patient respiration set.
This test is performed to inspect the patient breathing circuit. Any leakage in the patient
tubing system could be calculated using this test. In addition to the leakage, this test will also
measure total compliance of the patient tubing system.
Figure 7-3 System Test page
Press START button to start the system test. It should be noted that the circuit wye is closed
during this test. When the test is complete, select save button to store the calibration results.
59

Advertisement

Table of Contents
loading

Table of Contents