During the suction phase, flow value will be zero and the below alarms will be disabled for 2
minutes: Low PEEP, Low MV, Low VTe, Low Rate, APNEA, Disconnection
The message "Perform Suction and Reconnect within 120s" appears above the page during
the suction phase.
Final 100% O2
In this phase 100% O2 is delivered to the patient for maximum 2 minutes. Then the device
automatically returns to the previous settings and exits from the maneuver page.
If you press Stop button during this phase (2 minutes), the device will exit from this
maneuver and start working in the previous mode.
Note:
1- If the suction is carried out during the nebulization, the nebulizer will stop working.
2- Setting parameters or ventilation mode could not be changed during this maneuver.
3- Other menus and membrane keys are inactive during the Suction Support maneuver.
4- 1- This device is not intended to use closed loop suction. Therefore, the use of closed
loop suction may impair the performance of the device.

5-14 Shuting down the ventilator

To turn off the ventilator, standby the device and select the shut down key. The message
"Please wait..." will be displayed about 6 seconds. Then the ventilator will be shut down.
If it is not possible to shut down the device for any reason, you can turn off (hardware) the
ventilator by holding the On/Off button on the back of the device for about 10 seconds.
Figure 5-18 shuting down the ventilator