 Suction Support
The suction maneuver includes the below phases:
 Premier 100% O2
 Suction
 Final 100% O2
After selecting and starting each phase, the next phase will be enabled. Each phase is carried
out for maximum 2 minutes and then relevant button will be disabled and a progress bar is
filled up.
You can cancel the maneuver by pressing Stop button and you can exit from the Suction
Support page by pressing Back button. In the below conditions the suction support maneuver
will not be enabled and the message "SUCTION NOT Usable" will appear:
 When NIV is set On
 During 100% O2 maneuver
 When high pressure oxygen source is not available.
Premier 100% O2
At this stage, after pressing the Premier O2100% key, the device delivers breaths with 100%
oxygen for a maximum of 2 minutes. At this time, the user has 2 minutes to press the suction
key and disconnect the device from the patient for suction. If after 2 minutes of 100% oxygen
delivery time, the user does not press the suction button, the maneuver will end and the
system will automatically enter its previous breathing phase. In this case, the user must exit
the menu by pressing the Back key to activate the device menues.
The message "Disconnect Within 120 Seconds to Suction the Patient" appears above the
page during this phase.
Suction
After selecting suction, you must perform suctioning, connect the patient to the device and
press Final 100% O2 within maximum 2 minutes. If you press Final 100% O2 button
during this time, the device will enter final ventilation phase; otherwise the device will start
ventilation in the previous mode.