Operation
System test calibration
Air valve calibraion
Oxygen valve calibration
Exhalaion valve calibration
Periodic calibration
Air inlet filter check
Water and Humidity Drain
Check (Air Filter Glass
Enclosure)
Oxygen filter check
Oxygen sensor replacement
Exhalation flow sensor
replacement
Battery replacement
Exhalation valve- flow sensor
interface replacement
Exhalation valve membrane
replacement
Respiration set replacement
HME replcament
Exhalation filter replcament
Complete Antistatic Hoses
AIR (hospital Air to
compressor) replacement
Complete Antistatic Hoses
AIR (compressor to
ventilator) replace ment
Complete Antistatic Hoses O2
(Hospital to ventilator)
replacement
Humidifier chamber
replacement
Humidifier temperature probe
intervals
exhalation flow sensor or if
the expiratory volume is
incorrectly measured by the
sensor
Before using the ventilator
and after replacing the patient
breathing tube
Annually
Annually
Annually
Annually
Annually or every 5000 hours
of ventilator operation (each
occurred earlier)
Every 100 hours of ventilator
operation (4 days)
Annually or every 5000 hours
of ventilator operation (each
occurred earlier)
After the expiration date or
sensor failure
Annually or as needed
Every two years or during
battery failure
In case of damage
After each patient or in case
of damage
After each patient
After each patient, every 24
hours or as needed
After each patient, every 24
hours or as needed
In case of damage
In case of damage
In case of damage
After each patient
According to manufacturer
Responsible person
User
Expert
Expert
Expert
Expert
User
User
Expert
Expert
Expert
Expert
Expert
Expert
User
User
User
Expert
Expert
Expert
User
Expert
68