The maximum time taken to increase the oxygen concentration in the delivered gas to
the patient from 21% to 90% is 12 seconds (in the volumes 30, 150 and 300 ml).
The device maintenance should always be accomplished in conformity with safety
regulations.
The ventilator must be repaired, assembled, and used by trained personnel. The
ventilator must be inspected by trained personnel annually.
To prevent cross contamination, a bacteria filter can be placed between the ventilator
to patient outlet and the patient breathing circuit.
Do not reuse disposable filters and check them regularly for blockages, because
occlusion of these filters can cause high resistance of breathing circuit.
There will be some risks of the environment pollution associated with the disposal of
the device accessories and parts (e.g. battery, defective accessories). The device and
accessories shall be disposed of in compliance with relevant regulations. Contact your
municipality to check where you can safely dispose of old batteries.
Do not sterilize the ventilator.
If the ventilator is damaged, its life-supporting function will no longer be guaranteed.
In this condition do not use the ventilator and use other source of ventilation.
Do not use the ventilator placed next to a curtain, this could block cooling air thereby
causing the equipment to overheat.
To prevent electric shock while servicing the ventilator, ensure that it is disconnected
from all power sources.
Do not use the ventilator in magnetic resonance imaging (MRI) environment.
Do not use the ventilator in high- pressure enclosure.
Do not use the ventilator with nitric oxide.
Do not use the ventilator with helium or gas mixture containing helium.
Maintenance, sterilization and cleaning of the device should be conducted according
to the manufacturer recommendations in the user manual.
Use humidifier or HME to prevent delivery of dry and low temperature air to the
patient. Do not use the HME and the humidifier at the same time, because it increases
resistance of the airway.
Humidifier failure can cause delivery of low moisture and temperature air to the
patient.
Nebulizer or humidifier could increase resistance of breathing circuit filters. User
must check airway filters regularly.
To prevent possible damages, always turn off the humidifier when the ventilator is off
or in standby mode.
Nebulizer can add gas to the breathing circuit and affect the device measurement
accuracy.
In case of unavailability of the oxygen source, Nebulizer will not function.
To ensure that last settings and data are saved, shutdown the device by shutdown key
in standby menu.
Useful life of the device is 10 years. It is recommended to calibrate the device once a
year.