Saadat Respina P1 User Manual page 103

Table of Contents

Advertisement

Alarm Message
Apnea
High Leak
Plim Reached
High Oxygen
Low Oxygen
Volume Not
Delivered
Disconnection
Occlusion
High O2 Pressure
Low O2 Pressure
High Air Pressure
Possible Cause
No breath has been delivered in set
mode (no patient triggered breath or
delivered breath by the ventilator)
Leak value has exceeded alarm limit.
In volume based modes and during
inhalation if the airway pressure
exceeds" high Ppeak alarm -5"
The oxygen concentration is too high
compared with high alarm setting.
The oxygen concentration is too low
compared with low alarm setting.
The set volume has not been delivered
to the patient (in one of PRVC-CMV,
PRVC-SIMV or VSV modes that the
pressure is determined by the device).
The breathing circuit from the
ventilator to the patient is disconnected.
The breathing circuit or expiration
valve has been occluded.
The oxygen inlet pressure is more than
the specified level.
The oxygen inlet pressure is lower
than the specified level.
This alarm will not be enabled if the
oxygen concentration is set to 21%.
The air inlet pressure is more than the
specified level.
Corrective Action
- If Ftrig is used, ensure that
exhalation flow sensor is faultless
and calibrated.
- Check the patient condition.
- Check the device settings
(especially ventilation mode and
trigger value)
- Check the alarm limits.
- If Ftrig is used, ensure that
exhalation flow sensor is faultless
and calibrated.
- Check the patient airway.
- Check the alarm limits.
- Ensure that exhalation flow
sensor is faultless and calibrated.
- If the alarm persists, contact the
Customer Service department.
- Check the patient airway.
- Check the device settings.
- Check the pewssure alarm range.
- Ensure that the oxygen sensor is
faultless and calibrated.
- Check the air and oxygen
sources.
- Ensure that the oxygen sensor is
faultless and calibrated.
- Check the oxygen source.
- Check the inlet pressure oxygen
source concentration.
- Check the device settings.
- Check the alarm limits (high limit
of Ppeak).
- Check the patient airway.
- Check the patient condition.
- Check the patient airway.
- Check expiration valve.
- Check the oxygen source.
- If the alarm persists, contact the
Customer Service department.
- Check the oxygen source.
- If the alarm persists, contact the
Customer Service department.
- Check the air source.
- If the alarm persists, contact the
Customer Service department.
103